Chapter 5.
Cleaning & Maintain
-Device
(1)
Recommend that pour the liquid in the container before met the full line.
(2)
Disinfected with diluted medical alcohol to clean the device surface after use.
(3)
Disinfected with diluted medical alcohol to clean the accessories, such as suction container, connection
tube, and suction tube…etc.
(4)
Charge the device for 3hrs at least once per month.
(5)
The reusable collection canister is single-patient use and should be replaced every 1 to 2 months.
-
Bacteria Filter
(1)
The filter can’t be cleaned.
(2)
If the filter gets wet, or becomes contaminated or clogged, it must be replaced.
Chapter 6.
Troubleshooting
Problem
Possible Cause
Solutions
Pump runs,
but no vacuum
1. Tubing isn’t connected properly.
2. Tubing breaks or leaks.
3. Container do not assembly properly.
4. Container breaks or leaks.
1. Assembly the tube properly.
2. Contact your equipment provider for repair
Unit does not
working when
power is opened
1.
Switch broken.
2.
PCB board broken.
Contact your equipment provider for repair
Vacuum
delivered, but
gauge did not
display the
values
1.
Tubing isn’t connected properly.
2.
Gauge broken
Contact your equipment provider for repair
Battery will not
hold a charge.
1.
Charging IC broken
2.
Battery life expires
1. Ensure that unit is equipped with battery.
2. Verify that charge light turns on.
3. Check electrical connections during
charging.
4. Contact your equipment provider for repair
Low vacuum
1.
Tubing isn’t connected properly.
2.
Pump broken.
3.
Suction container full.
1. Assembly the tube properly.
2. Contact your
equipment provider for repair
Abnormal sound
when the device
working
1.
Pump broken
2.
Waste liquid or water
countercurrent to the machine
Contact your equipment provider for repair