AC Tool Supply RH350 User Manual Download Page 4

RH350-en-GB_V2.5   11/14 

4

Backlight Display 

Press the 

[M

in

/AV

RST

]

 and 

[MEM]

 buttons together to turn on the backlight. It will remain on for ten 

seconds. 

Automatic Power OFF 

With the meter off, press and hold down the 

[ON/OFF

SET

]

 

+

 

[REC

HOLD

]

 buttons for more than 1 

second to enter auto power off settings.  
Press the 

[MEM]

 button to select non-sleep (n) or 2, 5, 10, 20, 40 and 60 minutes autosleep.  

Press 

[T2

WB

]

 to confirm the selection. 

Manual Datalogging

Manual datalogging allows the user to store and recall up to 99 readings with the press of a button. 

1. Press the 

[MEM]

 button to manually store a single reading into memory. The display will flash 3

times and the memory location number will be displayed below the humidity reading.

2. To view stored readings, press and hold the 

[REC

HOLD

]

 button for 2 seconds and “RECALL” will

blink next to the memory location number on the display.

3. Press the 

[MEM]

 button to manually increment through the stored readings.

4. Press 

[REC

HOLD

]

 

+

 

[MEM]

 for more than 1 second to clear the memory.

5. Press 

[REC

HOLD

]

 for more than 1 second to return to normal operation.

Note:

 In RECALL mode, pressing the 

[M

in

/AV

RST

]

 button for min/max/avg readings will auto search 

the stored readings in memory and display the minimum, maximum and average values of the 
stored readings. 

Note

: A printer port is located on the side of the meter; however it is not functional on this model. 

Calibration 

The following verification and calibration procedures require the 33% and 75% RH reference bottles 
which are supplied with the Model RH350-CAL kit. 

Accuracy Verification 

Checking the 33% or 75% RH Calibration 

1.

Insert meter’s sensor into the 33% or 75% salt reference bottle

2.

Check the reading after a minimum of 10 minutes

3.

Verify that the reading is within the accuracy specification

Relative Humidity Calibration (33% and 75%) 

33% Calibration 

1.

Turn the meter Off.

2.

Insert the meter sensor into the 33% salt reference bottle.

3.

Allow the meter one hour to stabilize.

4.

While holding down the 

[M

in

/AV

RST

]

 button, press and hold the 

[ON/OFF

SET

]

 button for 2

seconds to enter calibration mode.

5.

32.8%

” (reading may vary depending on operating temperature) will flash on the LCD

screen. After 30 minutes, the flashing will stop to indicate that the 33% calibration is
complete.

Note: Do NOT turn off the meter 

6.

Proceed to the 75% Calibration below.

actoolsupply.com

actoolsupply.com