Pentra
80
4–54
Pentra
80 - User Manual - RAB108EA
6.2.2. Association grid keys
Tab. 4–12: Association function Keys
6.2.3. Results/orders matching
This screen will allow the user to manually match Sample tubes and Worklist Orders that have
been classified as «Exceptions» (see
Select a result (
in the lower half of the association grid
) that you want to associate an order
with.
Now select the order (
in the upper half of the association grid
) to associate to the selected
result.
Then select the «OK» key. A confirmation message will be displayed (see
Fig. 4–45 Matching confirmation
Once the matching has be performed, the selected lines will disappear. The matched results
will be added to the «Result» grid (see
6.1.3. Result Grid information,
A notification is added to the instrument logs.
◆
If the order type is different from the results one, the results will be recomputed
(with the new type settings: thresholds, limits....). The recomputed results will
be flagged in the «Remarks» flags (see
◆
If the order test differs from the results test, then the test of the matched results
= CBC
◆
When the user validates the association and no order is selected, the following
message will be displayed: «
Do you still want to match the tube with its default
order?
».
Heading / Key
Name
Function
Delete
Deletes selected order of the Worklist: a confirmation
message is displayed «
Do you want to delete the order
which sample ID is XXX
»
OK
Associates the selected order with the selected results
(see
6.2.3. Results/orders matching,
notification is done in the logs.
Click a line
Click a line
On the result list:
- Displays the Rack and position of the tube; displays
the running date too.
- Allows to match the tube with its default order (if no
order is selected)
Summary of Contents for Pentra 80 RAB108BA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE ...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11 ...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options ...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab ...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA ...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA ...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA ...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28 ...