PROBLEM: 

POSSIBLE CAUSE / SOULTIONS: 

Moisture collects inside 

 

Door gasket is not sealing properly. Check for debris, 
cracks, and items passing through door at the gasket.                                                             

 

The refrigerator or freezer is facing a doorway or is 
underneath of air conditioning vent. Relocate the unit or 
redirect air vent. 

 

Too many door openings. Minimize time door is open. 

 

Hot, humid weather increases condensation. 

 

Make sure there is a water trap (U-shaped loop) in the 
drain tube near the compressor. This will “trap” a small 
amount of water in the loop and prevent air from 
entering the chamber through the tube.  

Moisture collects on outside 
surface 

 

Hot, humid weather increases condensation. 

 

As humidity decreases, moisture will disappear. 

The display shows “SO” and 
the compressor does not run 

 

The temperature sensor is open. Call for service.                                                                                                                                                                                                                                                                                                                                                       

The display shows “SC” and 
the compressor does not run 

 

The temperature sensor is shorted. Call for service.  

The display shows “HH” and 
the compressor does not run 

 

The ambient temperature is above 40C/104F. Stop using 
immediately. 

The display shows “LL” and the 
compressor does not run 

 

The ambient temperature is below -40C/-40F. Stop using 
immediately.  

MOISTURE DURING THE SUMMER SEASON 

The amount of moisture, condensation, or high humidity related issues increase during the summer 
and, in most cases, will self-resolve when the weather cools down.  Please note a refrigeration 
system does NOT generate moisture or water but simply condenses the moisture that is already in 
the chamber. Keeping the unit in an air conditioned, low humidity space will resolve many issues. 
Other things you should check 

1.

 

Location of the refrigerator (See Quick Troubleshooting Guide above) 

2.

 

Door sealing and frequency of door opening event (See Quick Troubleshooting Guide above) 

3.

 

Make sure there is a water trap (U-shaped loop) in the drain tube near the end. This will 
“trap” a small amount of water in the loop and prevent air from entering the chamber 
through the tube.  

 
BEFORE CALLING THE MANUFACTURER’S TECHNICAL SUPPORT DEPARTMENT, 

please have the 

unit’s model and serial number ready as well as the problem description.  The model and serial 
number is located on the serial tag which can be found on the interior left upper wall of the 
unit. 

Summary of Contents for ABT-HC-UCBI-0420-ADA

Page 1: ...me of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is being unpacked Whe...

Page 2: ...SERVICE please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left OR righ...

Page 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Page 4: ...is cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection VOLTAGE SHOULD NOT VARY M...

Page 5: ...ecially if the weather is very hot This is normal The motor compressor will cycle normally as soon as the excess heat has been removed TEMPERATURE CONTROL NOTE The control is preset at the factory to...

Page 6: ...s of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inches...

Page 7: ...ides the following functions Red LED for visual indication of an alarm condition Beeper for audible indication of an alarm condition A preprogrammed 5 minute delay from the time a set point is attaine...

Page 8: ...Pon The control keys are now unlocked and you can perform the programming functions below Adjusting the temperature control set points NOTE The temperature control set points are preset at the factor...

Page 9: ...isplay Press the UP or DOWN keys several times to cycle through the parameters until CF is displayed Press SET to display the current setting either C for Celsius or F for Fahrenheit Use the UP or DOW...

Page 10: ...d the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or side walls Evaporator is blocked by fros...

Page 11: ...bove 40C 104F Stop using immediately The display shows LL and the compressor does not run The ambient temperature is below 40C 40F Stop using immediately MOISTURE DURING THE SUMMER SEASON The amount o...

Page 12: ...not use commercial household cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smooth handles Do not use a dry cloth to clean smooth handles STAINLESS STEEL CARE AND CLEANING Use...

Page 13: ...removal on all finishes Grease oil Any good commercial detergent Apply with a sponge or cloth Restoration Passivation Benefit Super Sheen Passive Film Breakdown Corrosion on metals is everywhere We r...

Page 14: ...ging to avoid electrical hazard DO NOT ADJUST THE TEMPERATURE CONTROL Remove the product and leave the door open while defrosting the freezer Place towels in the bottom of the freezer to catch the fro...

Page 15: ...e paid by the customer ADDITIONAL THREE YEAR COMPRESSOR WARRANTY In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compr...

Page 16: ...t damage during transit or installation fire flood or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with t...

Reviews: