background image

 

Copyright 

 2006, Delmhorst Instrument Co.  

Page 21 of 32 

established with the base unit. The base unit will also indicate communication 
established by turning off the green LED and turning on a red LED.  
 

3. 

Press the “MANAGE DATA” button in the Delmhorst Instrument Co.

 

Universal Application Program. The meter will display “

Data

” on the screen 

and the base unit will alternately turn on a yellow and red LED while data 
transfer takes place (red indicates communication between the meter and 
Base Unit, while yellow indicates communication between the base unit and 
the PC). When finished, the meter will return to the Main Menu and the base 
unit will, once again, turn on its green LED. For further details on managing 
the meter data in the PC please refer to the HELP feature in the program. 

 

 

NOTE: 

The above sequence of events in initiating communication between 

the PC and the meter is not mandatory. The Application Program may be 
started first, or the IR Link may be selected first. The 

MANAGE DATA

 button 

may also be pressed before communication is started between the Base Unit 
and the meter. The meter, base unit and Application Program all have ample 
delays built in to allow the other link components to come online in any 
sequence. 
 

 

TROUBLESHOOTING THE IR LINK

Most problems tend to occur due to the positioning of the meter relative to the 
base unit. There are a few error messages that may appear on the meter’s 
screen: 

 

a. 

Timeout

: This error message occurs when too much time has passed 

between data sent by the meter and the response that should have been 
received from the base unit. Possible causes: 

 

  Meter misalignment with the Base Unit (make sure the communication 

windows are facing each other, within the distance indicated). 

 

  No connection to the PC. The Base Unit must be connected to the PC 

through a USB cable. Check connection. 

 

b. 

KBD:

 This message occurs when the user presses a key during 

communications. Do not push any key on the meter during data transfer. The 
meter interprets this as a request to escape this activity and returns to the 
Main Menu. The base unit should return to a green LED

 

status in a few 

seconds. A message should appear on the PC screen as well, indicating lost 
communications. 

 
c. 

Sync:

 This message occurs when data framing is lost during data 

transmission. This would only happen if the wireless link becomes unreliable. 
Ensure the meter and base unit are aligned and restart the session. 

d. 

Data E

: This message occurs when the content of a correctly framed data 

packet is invalid. Again, this is an unreliable wireless link problem. 

 

Summary of Contents for TotalCheck

Page 1: ...Version 3 02 For Total Check Meters starting with Serial 12468 Delmhorst Instrument Co Rev 13 June 11 2014 Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web...

Page 2: ...9 To Set the Alarm in Basic Scan Mode 10 To Take TEMP RH Readings in Basic Mode 10 To Reset the Meter to Default Settings in Basic Mode 11 USING THE METER IN ENHANCED MODE 11 Job Definition 11 To Cre...

Page 3: ...esting Wood 24 Paint Failure Moisture 25 EIFS 25 Using the 0 100 Numerical Reference Scale 26 Testing Concrete Slabs for Flooring Applications 26 Measuring RH in Concrete Slabs 27 Testing Insulation 2...

Page 4: ...ilding materials Scan Mode o Reference scale 0 300 Thermo hygrometer o RH Sensor includes RH T S3for ASTM F 2170 in situ testing of concrete slabs RH T S1 available for general industrial ambient appl...

Page 5: ...te the selection to any other available option The keypad is aimed at providing navigational control and not at accessing specific features There are four directional keys aligned intuitively around a...

Page 6: ...the download is complete perform a meter reset by choosing RESTORE DEFAULTS in the Storage Menu If a download is not necessary go directly to RESTORE DEFAULTS LOW BATTERY The meter features a battery...

Page 7: ...et up a job definition BASIC mode does not allow for data collection and is the default mode of the meter upon power up Or TotalCheck may be used in ENHANCED mode which requires the user to set job de...

Page 8: ...erting the pins into the material and pressing the SELECT key This will display a MC reading If you wish to hold a reading on screen press the SELECT key again otherwise remove the pins from the mater...

Page 9: ...the meter will make the correction 1 To change temperature use the keys from the MC READ screen to highlight the TEMPERATURE T 70 F Press the SELECT button to get into the TEMPERATURE sub menu 2 To sc...

Page 10: ...OFF Press the SELECT key to turn it on 3 After the alarm is turned on use the keys to scroll up or down to a higher or lower alarm value 4 After you have selected the alarm value use the key to highl...

Page 11: ...meter will return to factory default settings in basic mode i e Species Douglas Fir Temperature correction 70 F Alarm OFF in pin mode In Scan mode Alarm off USING THE METER IN ENHANCED MODE To use Tot...

Page 12: ...a new job the Create Job screen will appear as follows At the Create Job screen you may select the material being tested specify the temperature at which measurements will be taken and select an alarm...

Page 13: ...MP TEMP sub menu 3 Using the keys scroll up or down to a higher or lower temperature value Temperature values will change in 5 F increments 4 To change between C and F press the SELECT key when the CO...

Page 14: ...e SELECT key to toggle between TRIG and CONT When the meter is set to Triggered mode a single reading is taken every time the SELECT button is pressed Each reading is stored in memory The last measure...

Page 15: ...main menu This will bring you to the SCAN screen 2 Use the and keys and highlight ALARM OFF Press the SELECT key to enter the alarm sub menu 3 The default status of the alarm is OFF Press the SELECT k...

Page 16: ...to highlight your selection Press the SELECT key to choose that variable MAIN MENU OPTIONS STORAGE The Storage Menu groups all meter features related to measurement storage retrieval and manipulation...

Page 17: ...eter will display a screen with the word PROCESSING while calculations take place User input is ignored during this time Control is returned to the user once the algorithm is complete and the second s...

Page 18: ...the SELECT key to choose your selection 3 By using the keys you may select which job and location you wish to view If there are no readings in a job the following screen will display If there area rea...

Page 19: ...e Left button causes the meter to subtract 10 from the current left side value o DONE Selecting this option signals to the meter that the user has completed the reading recall setup process and is rea...

Page 20: ...he red side window of the meter and the receiving window of the base unit Simply place the meter and base unit on a flat surface with the red side window of the meter facing the receiving window of th...

Page 21: ...er link components to come online in any sequence TROUBLESHOOTING THE IR LINK Most problems tend to occur due to the positioning of the meter relative to the base unit There are a few error messages t...

Page 22: ...Time This menu option allows user access to the date and time set screen The Date and Time set screen is populated with the time and date values present in the meter at access time key action switche...

Page 23: ...abled Backlight shutdown delays are 10 20 30sec It is recommended that the backlight delay be set to the minimum amount of time necessary in order to maximize battery life Conversions This menu option...

Page 24: ...grain and push them to their full penetration into the wood if possible Insulated pins read only at the tip and can be driven to the desired depth Press the SELECT button and read the moisture conten...

Page 25: ...in contact with the ground may absorb moisture Siding and shingles without sufficient lap so that water is forced up through cracks by wind pressure Ice dams Condensation of vapor within hollow walls...

Page 26: ...e TESTING CONCRETE SLABS FOR FLOORING APPLICATIONS Moisture meters are an effective tool to check comparative moisture conditions in concrete slabs They can tell you where there may be excess moisture...

Page 27: ...oisture content in Concrete For example readings in the 68 to 77 range indicate approximately 2 to 4 moisture content When evaluating a slab for readiness always consider its age thickness whether the...

Page 28: ...t www ASTM org or contact Delmhorst customer service Refer to the White Paper Relative Humidity Sensor Behavior and Care on our website for additional information regarding the sensor itself TESTING I...

Page 29: ...ton is pressed and released again during the reading cycle Note Readings between 0 1 and 0 5 indicate a sufficiently dry moisture level Readings between 0 5 and 1 indicate a borderline situation Readi...

Page 30: ...for minimum 1 2 hours or as long as possible Clean the meter with any biodegradable cleaner Use the cleaner sparingly and on external parts only Keep cleaner out of the external connector Remove the b...

Page 31: ...uld stay in calibration follow the instructions in the Care of Your Meter section UNDER NO CIRCUMSTANCES SHALL DELMHORST BE LIABLE FOR ANY INCIDENTAL INDIRECT SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY T...

Page 32: ...Y 54 PINE SOUTHERN YELLOW 19 COCOBOLO 55 PINE SUGAR 20 COTTONWOOD 56 PINE WHITE 21 CUMARU 57 POPLAR YELLOW 22 CYPRESS 58 PURPLEHEART 23 EBONY AFRICAN 59 RAMIN 24 ELM AMERICAN 60 REDWOOD 25 FIR RED 61...

Reviews: