Abiomed AB5000 Quick Start Manual Download Page 6

0015-7309, RevH

 

 AB5000™ System Installation / Preventative Maintenance Procedure  

Page 6 of 8

 

ABIOMED, Inc.  Proprietary and Confidential 

SOP-QA1000-T2, Rev. B 

 

[ ]   24.  TARE OFFSETS 

Connect diagnostic download cable between Computer and Console. Power Console on. Enter 
Diagnostic Terminal mode.  Record initial tare offset values at ‘0 min (a)’.  

If the tare value is outside of +/- 10.0 L/min, replace the flowmeter.  

If the tare value is outside of  +/- (3.0 – 10.0 L/min) , send the following commands to tare the 
flowmeter (items within <> indicate the key to press): 

 

 

<

ENTER> <ENTER> *@=A <ENTER> A$$V <ENTER> 

The tare offset should be reduced following the command. Repeat once more if it does not. 
Record the post-tare offset into 0(b).  

If the tare value is /- 3.0 L/min, record the offset into 0 (a), record n/a into 0(b). 

Record the tare offset in the following time intervals. 

 

Left Alicat 

Right Alicat 

Time 

(min) 

Flow LPM 

Flow LPM 

Acceptable Flow Out: 

Pass 

 Fail 

 

0 (a) 

 

 

+/- ( 3.0 – 10.0 ) L/min 

N/A 

N/A 

0 (b) 

 

 

0.0 +/- 3.0 L/min 

 

 

 

 

T=2-T=0 < 

±

0.5 L/min

 

 

 

 

 

T=4-T=2 < 

±

0.5 L/min

 

 

 

 

 

T=6-T=4 < 

±

0.5 L/min

 

 

 

 

 

T=8-T=6 < 

±

0.5 L/min

 

 

 

10 

 

 

T=10-T=8 < 

±

0.5 L/min

 

 

 

12 

 

 

T=12-T=10 < 

±

0.5 L/min

 

 

 

14 

 

 

T=14-T=0< 

±

0.5 L/min

 

 

 

16 

 

 

T=16-T=14 < 

±

0.5 L/min

 

 

 

18 

 

 

T=18-T=16 < 

±

0.5 L/min

 

 

 

20 

 

 

T=20-T=18 < 

±

0.5 L/min

 

 

 

22 

 

 

T=22-T=20 < 

±

0.5 L/min 

 

 

24 

 

 

T=24-T=22 < 

±

0.5 L/min 

 

 

26 

 

 

T=26-T=24 < 

±

0.5 L/min 

 

 

28 

 

 

T=28-T=26 < 

±

0.5 L/min 

 

 

30 

 

 

T=30-T=16< 

±

0.5 L/min 

 

 

 

Specification: 0.0 +/- 3.0 L/min (excluding 0 (a)) 

 

 

Summary of Contents for AB5000

Page 1: ...ater Serial Port Diagnostic Download Cable ABIOMED IMPELLA P N 0015 9612 Pump ID Connector ABIOMED IMPELLA P N 0055 0737 Pressure Vacuum Gauge 200mmHg to 500mmHg min Damping set to 1 sec Safety Analyz...

Page 2: ...andle is tight Move the left and right shuttle switches to BVS Pump mode Verify shuttle pins engage securely into ball plungers Cycle the Hand Pump Handle multiple times Check for airflow at the left...

Page 3: ...the left OFF LED blinks red The left OFF LED turns off in 10 14 seconds Press the left OFF button twice within 10 seconds Verify The Console stops pumping the left OFF LED illuminates red the left ON...

Page 4: ...de LOW PRESSURE LOW FLOW alarms occur Press the right OFF button twice Block the right barb then press the right ON button Verify Within 9 beats the right side HIGH PRESSURE LOW FLOW alarms occur Pres...

Page 5: ...mp ON Ventricle Mode Pressure Pump OFF Ventricle Mode Vacuum Pump OFF Left 320mmHg 15 35mmHg 10 420mmHg 15 100mmHg 10 Right 200mmHg 10 35mmHg 10 300mmHg 10 100mmHg 10 NOTE Pressure Values to be measur...

Page 6: ...re offset should be reduced following the command Repeat once more if it does not Record the post tare offset into 0 b If the tare value is within 3 0 L min record the offset into 0 a record n a into...

Page 7: ...plug to the remote alarm jack Connect a DVM between the plug contacts Verify Short circuit Ohmic resistance of 1 ohms when alarm condition is active Open circuit Ohmic resistance more than 1M is acce...

Page 8: ...ect castors for any damage Check brake directional locks and rotation Check storage compartment door for proper operation and latching Check Console to Cart latching mechanism for proper operation 30...

Reviews: