0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure
Page 6 of 8
ABIOMED, Inc. Proprietary and Confidential
SOP-QA1000-T2, Rev. B
[ ] 24. TARE OFFSETS
Connect diagnostic download cable between Computer and Console. Power Console on. Enter
Diagnostic Terminal mode. Record initial tare offset values at ‘0 min (a)’.
If the tare value is outside of +/- 10.0 L/min, replace the flowmeter.
If the tare value is outside of +/- (3.0 – 10.0 L/min) , send the following commands to tare the
flowmeter (items within <> indicate the key to press):
<
ENTER> <ENTER> *@=A <ENTER> A$$V <ENTER>
The tare offset should be reduced following the command. Repeat once more if it does not.
Record the post-tare offset into 0(b).
If the tare value is /- 3.0 L/min, record the offset into 0 (a), record n/a into 0(b).
Record the tare offset in the following time intervals.
Left Alicat
Right Alicat
Time
(min)
Flow LPM
Flow LPM
Acceptable Flow Out:
Pass
√
Fail
√
0 (a)
+/- ( 3.0 – 10.0 ) L/min
N/A
N/A
0 (b)
0.0 +/- 3.0 L/min
2
T=2-T=0 <
±
0.5 L/min
4
T=4-T=2 <
±
0.5 L/min
6
T=6-T=4 <
±
0.5 L/min
8
T=8-T=6 <
±
0.5 L/min
10
T=10-T=8 <
±
0.5 L/min
12
T=12-T=10 <
±
0.5 L/min
14
T=14-T=0<
±
0.5 L/min
16
T=16-T=14 <
±
0.5 L/min
18
T=18-T=16 <
±
0.5 L/min
20
T=20-T=18 <
±
0.5 L/min
22
T=22-T=20 <
±
0.5 L/min
24
T=24-T=22 <
±
0.5 L/min
26
T=26-T=24 <
±
0.5 L/min
28
T=28-T=26 <
±
0.5 L/min
30
T=30-T=16<
±
0.5 L/min
Specification: 0.0 +/- 3.0 L/min (excluding 0 (a))