 AB5000™ System Installation / Preventative Maintenance Procedure  

Page 2 of 8

 

ABIOMED, Inc.  Proprietary and Confidential 

SOP-QA1000-T2, Rev. B 

 

[ ]   6.  CHECK TIME METER HOURS.  REPLACE PARTS AS REQUIRED PER MAINTENANCE SCHEDULE

 

0015-3050 Humphrey® Valve Assembly 

Every 5,000 Hours 

0015-3000 Proportional Valve Assembly 

Every 5,000 Hours 

0015-3040 Compressor Assembly  

                          Every 5,000 Hours 

0015-3060 Battery Module Assembly 

                          Every 2 Years 

0015-3061P Battery Module Assembly A1 

             Every 2 Years 

0015-7309 Perform PM Procedure 

                          Annually 

 

[ ]   7.  CONSOLE LUGGAGE HANDLE 

[ ]

 

Check Console’s luggage handle for smooth operation. Verify proper latching in raised 
position. 

[ ]

 

Check the console’s luggage handle for chips, cracks, or signs of abuse. Replace if 
necessary.  

[ ]

 

If available, check Bed Plate assembly for socket head cap crew, replace with flat 
countersunk head cap screw (PN 2850-5127). 

[ ]   8.  HAND PUMP 

[ ] 

Inspect the Hand Pump for any physical damage or loose hardware. Verify Piston Rod 
Handle is tight.

 

Move the left and right shuttle switches to BVS Pump mode. 

[ ]

 

Verify shuttle pins engage securely into ball plungers.

 

[ ] 

Cycle the Hand Pump Handle multiple times. Check for airflow at the left and right barbs.

 

Move the left and right shuttle switches to Ventricle Pump mode.

 

[ ] 

Verify shuttle pins engage securely into ball plungers.

 

[ ] 

Cycle the Hand Pump Handle multiple times. Check for airflow at the left and right barbs.

 

[ ] 

Inspect Console’s Hand Pump mount for any physical damage.  Place the Hand Pump 
back into the mounting bracket.

 

 

[ ]   9.  POWER-UP SELF TEST SEQUENCE 

Plug the Console IN. Turn the Power Switch ON. Verify: 

[ ] 

All Keypad LED indicators illuminate and that the audible alarm sounds.

 

[ ] 

An air pulse at the left barb, then an air pulse at the right barb. 

 

[ ] 

The self test ends within twenty (20) seconds and then the LCD displays the message “LEFT 
SYSTEM READY FOR USE” & “RIGHT SYSTEM READY FOR USE”. 

 

[ ]   10.  FAN CHECK 

[ ]

 

Verify the operation of the system cooling fan (access at the Diagnostic Access Panel). 

 

[ ]   11.  POWER SUPPLY 

Setup: Console plugged in, batteries connected, console power on. 

 

[ ] 

Measure voltage output in the Power Supply assembly. Verify that voltage is 32 ± 0.1 VDC.

  

 

 

 

Summary of Contents for AB5000

Page 1: ...ater Serial Port Diagnostic Download Cable ABIOMED IMPELLA P N 0015 9612 Pump ID Connector ABIOMED IMPELLA P N 0055 0737 Pressure Vacuum Gauge 200mmHg to 500mmHg min Damping set to 1 sec Safety Analyz...

Page 2: ...andle is tight Move the left and right shuttle switches to BVS Pump mode Verify shuttle pins engage securely into ball plungers Cycle the Hand Pump Handle multiple times Check for airflow at the left...

Page 3: ...the left OFF LED blinks red The left OFF LED turns off in 10 14 seconds Press the left OFF button twice within 10 seconds Verify The Console stops pumping the left OFF LED illuminates red the left ON...

Page 4: ...de LOW PRESSURE LOW FLOW alarms occur Press the right OFF button twice Block the right barb then press the right ON button Verify Within 9 beats the right side HIGH PRESSURE LOW FLOW alarms occur Pres...

Page 5: ...mp ON Ventricle Mode Pressure Pump OFF Ventricle Mode Vacuum Pump OFF Left 320mmHg 15 35mmHg 10 420mmHg 15 100mmHg 10 Right 200mmHg 10 35mmHg 10 300mmHg 10 100mmHg 10 NOTE Pressure Values to be measur...

Page 6: ...re offset should be reduced following the command Repeat once more if it does not Record the post tare offset into 0 b If the tare value is within 3 0 L min record the offset into 0 a record n a into...

Page 7: ...plug to the remote alarm jack Connect a DVM between the plug contacts Verify Short circuit Ohmic resistance of 1 ohms when alarm condition is active Open circuit Ohmic resistance more than 1M is acce...

Page 8: ...ect castors for any damage Check brake directional locks and rotation Check storage compartment door for proper operation and latching Check Console to Cart latching mechanism for proper operation 30...