 AB5000™ System Installation / Preventative Maintenance Procedure  

Page 1 of 8

 

ABIOMED, Inc.  Proprietary and Confidential 

SOP-QA1000-T2, Rev. B 

 

 

AB5000™  System Console 

S/N 

AB5000™  AB Cart OR 
iPulse Cart 

S/N 

Hand Pump 

S/N 

Time Meter 

Hrs. 

Date  
Tested By 

 

 
All service and maintenance, which involves disassembly of the AB5000

TM

 Console, must be performed by factory 

trained and certified service personnel.  Replacement parts obtained from sources other than ABIOMED/IMPELLA 
must not be used. Possible  Electrostatic discharge precautions must be observed. Prior to servicing, check with 
ABIOMED/IMPELLA technical support (USA 800-422-8666) (49-241 8860301) for available FCN 
releases and Service Manual updates. 

 

REQUIRED EQUIPMENT 

 

Item 

Specification (or ABIOMED P/N for Specialized 
Tools) 

Angled mirror 

Miltex Mirror P/N 67-696/3 and Mirror Handle P/N 011-
26641  or equivalent 

Lint free swab with 6“ plastic handle 

Abiomed PN 7000-0030 (Lint free swab with 6” plastic 
handle) or equivalent 

Hydraulic Load Simulator 

ABIOMED/IMPELLA P/N 0005-0070 

Digital Voltmeter 

Fluke 77 or Equivalent 

Laptop Computer 

Running Windows 2000 or Later, Serial Port 

Diagnostic Download Cable 

ABIOMED/IMPELLA P/N 0015-9612 

Pump ID Connector 

ABIOMED/IMPELLA P/N 0055-0737 

Pressure/Vacuum Gauge 

-200mmHg to +500mmHg min (Damping set to 1 sec.) 

Safety Analyzer (EUROPE ONLY) 

IEC601-1 Tester 

 

[ ]   1.  DECONTAMINATION

 

Prior to servicing, the AB5000

TM 

Console must be decontaminated. Unless otherwise known, the Console 

should be considered contaminated and universal safety precautions are to be taken. Gloves, masks, eye 
protection, and protective garments are to be worn as required in the handling and decontamination of the 
Console. 
 

[ ]   2.  UNCRATING

 

Uncrate the Console and Cart.  Place the Hand Pump into the mounting bracket as required. 

[ ]   3.  EXTERNAL INSPECTION

 

Perform a thorough

 

exterior inspection of the Console and Cart. Inspect for: 

[ ] 

Turret rotation, panel damage, mechanical damage and any loose hardware. 

[ ]   4.  INTERNAL INSPECTION

 

Remove Console from Cart. Remove the Console panels. Perform a thorough

 

interior visual inspection. 

(Inspect fittings, tubings, quick-connect tubing (insertion depth), cable connections, hardware). 

[ ]

 

Perform a general inspection of the compressor subassemblies, including but not restricted to, 

black debris in the rear bearing, tubing wear, mesh wear and electrical connections as per procedure 
0015-7007. 

[ ]   5.  PERFORM REQUIRED FIELD CHANGES (FCNS) – CONTACT ABIOMED FOR CURRENT FCN 

STATUS

 

 

Summary of Contents for AB5000

Page 1: ...ater Serial Port Diagnostic Download Cable ABIOMED IMPELLA P N 0015 9612 Pump ID Connector ABIOMED IMPELLA P N 0055 0737 Pressure Vacuum Gauge 200mmHg to 500mmHg min Damping set to 1 sec Safety Analyz...

Page 2: ...andle is tight Move the left and right shuttle switches to BVS Pump mode Verify shuttle pins engage securely into ball plungers Cycle the Hand Pump Handle multiple times Check for airflow at the left...

Page 3: ...the left OFF LED blinks red The left OFF LED turns off in 10 14 seconds Press the left OFF button twice within 10 seconds Verify The Console stops pumping the left OFF LED illuminates red the left ON...

Page 4: ...de LOW PRESSURE LOW FLOW alarms occur Press the right OFF button twice Block the right barb then press the right ON button Verify Within 9 beats the right side HIGH PRESSURE LOW FLOW alarms occur Pres...

Page 5: ...mp ON Ventricle Mode Pressure Pump OFF Ventricle Mode Vacuum Pump OFF Left 320mmHg 15 35mmHg 10 420mmHg 15 100mmHg 10 Right 200mmHg 10 35mmHg 10 300mmHg 10 100mmHg 10 NOTE Pressure Values to be measur...

Page 6: ...re offset should be reduced following the command Repeat once more if it does not Record the post tare offset into 0 b If the tare value is within 3 0 L min record the offset into 0 a record n a into...

Page 7: ...plug to the remote alarm jack Connect a DVM between the plug contacts Verify Short circuit Ohmic resistance of 1 ohms when alarm condition is active Open circuit Ohmic resistance more than 1M is acce...

Page 8: ...ect castors for any damage Check brake directional locks and rotation Check storage compartment door for proper operation and latching Check Console to Cart latching mechanism for proper operation 30...