15
9. Connectors and cables
Connector symbols can be found at the bottom of the
control unit. (picture 3)
v
X1
Connector for an external power supply. Only
use Globtek inc. power supply (see chapter
Technical Specifications
for details) which can
be obtained as an original accessory from Abilia.
X2
AUX Connector to connect the device to a nurse call system, a personal
emergency phone or an external wireless transmitter. The connector may only be
connected to a system safety voltage input with max. voltage below 25 V(AC) / 60 V (DC),
where both poles have been separated from the electrical network. Max. load current
100 mA.
X3
- Sensor connector. Only use the Emfit bed sensor.
Connect the bed sensor (picture 13), the power supply delivered as an accessory (picture
14) and any connector cables (picture 15) according to the pictures.
9.1 X2 (AUX) Connector Pin Order (picture 16)
From left to right:
Pin 1
Common return
Pin 2
Normally Open (NO) send
Pin 3
Normally Closed (NC) send
Pin 4
Low Battery send (NC)
Pin 5
Not in use, do not connect
Pin 6
Not in use, do not connect
Pin 7
Not in use, do not connect
Pin 8
Not in use, do not connect
10. Installation of the control unit
10.1 With wall mounting bracket
1) Fix the mounting bracket onto a wall with the plugs and screws supplied.
(picture 17)
2) Slide the control unit onto the wall mounting bracket.
(picture 18)
3) Press the control unit down until you hear a click.
(picture 19)
10.2 With bed side clip
Press the control unit down to the clip.
(picture 20)
10.3 On table
Place the control unit on the table led lights facing upwards.