Abilia Emfit Epilepsy Alarm Manual Download Page 1

 

 

Rev F1 EN 

 

 

 

 

 

 

 

 

Emfit

®

 Tonic-Clonic Seizure Monitor 

 

 

 

Art.no 464000, 464031 

 

 

 

 

 

Note! The device must not be used if an operational failure of the device might lead to a 

delay in getting the appropriate treatment or medication for the patient, which could put the 

patients life at risk. 

M a n u al  

Summary of Contents for Emfit Epilepsy Alarm

Page 1: ...onitor Art no 464000 464031 Note The device must not be used if an operational failure of the device might lead to a delay in getting the appropriate treatment or medication for the patient which coul...

Page 2: ...2...

Page 3: ...6 3 SW 1 switch function switch 6 12 6 4 Adjusting the notification sound volume switches 7 and 8 12 7 Inserting batteries and battery service life 13 8 5V AC External power supply optional 14 9 Conne...

Page 4: ...al of the device after use 23 21 Declaration of conformity EU 23 22 Abilia limited warranty statement 23 22 1 Your guarantee 23 22 2 Terms 23 22 3 Exclusions and limitations 24 22 4 Your legal rights...

Page 5: ...nications or local area networks If this device is used with a pressure care mattress filled using a compression pump the device may not function normally in some cases Equipment is not suitable for u...

Page 6: ...se chapter Electromagnetic conditions Symbol of European Waste Electrical and Electronic Equipment Directive WEEE Directive 2002 96 EC on waste electrical and electronic equipment in these instruction...

Page 7: ...certified in the external power supply Product is C TICK certified in the external power supply Product is China RoHS 30 certified in the external power supply Product is BSMI certified in the externa...

Page 8: ...consists of a bed sensor located under the mattress and a control unit where the bed sensor is connected Control unit should be placed on a table clipped on bed or mounted on wall The control unit no...

Page 9: ...operator is the caregiver of the user The operator needs to be adult who fully understands instructions and limitations of use mentioned in these instructions for use Otherwise no special skills or k...

Page 10: ...details picture 2 In the event of a power failure two high quality 1 5V AA alkaline batteries can be used as an emergency power supply Do not use any rechargeable batteries or lithiumion batteries Re...

Page 11: ...er supply and remove nurse call cable if used from connector X2 before opening the lid Remove the batteries and disconnect the power supply before setting up the DIP switches in order to implement the...

Page 12: ...switched on or off by pressing the switch for three 3 seconds OFF down The SW1 switch does not act as the On Off switch The device is always on when it is connected to the power supply or when the bat...

Page 13: ...is someone in bed 75 of that time there are two notifications per day and sound notification is on 30 seconds each time When the batteries are getting low the red LED will begin to flash slowly A beep...

Page 14: ...s connected the batteries act as backup power supply in the event of a power failure All alkaline batteries self discharge and start to leak when empty contaminating the device Ensure that batteries a...

Page 15: ...the bed sensor picture 13 the power supply delivered as an accessory picture 14 and any connector cables picture 15 according to the pictures 9 1 X2 AUX Connector Pin Order picture 16 From left to rig...

Page 16: ...place the bed sensor under the mattress or mattress topper never just under the sheet The bed sensor must not come into direct contact with a person The bed sensor is designed to last for a minimum o...

Page 17: ...press the switch for too long you might accidentally switch off the device The notification sound will also stop when the fast paced movements stop 12 3Bed exit alarm bypass switch The SW1 switch act...

Page 18: ...f No one is on the bed 13 2Blue Device on standby Light is off Device is switched off Light flashes slowly Device is switched on Light flashes rapidly a few times Device triggers a notification 13 3Re...

Page 19: ...re high priority signals that will overrule information signals if both signals activate simultaneously 14 1Alarm signals Two high and low BEEPS in a loop Fast movements alarm goes off Three high and...

Page 20: ...bed should not last long enough for the notification to sound Person weight Switch position in parenthesis if false alarms occur 75 kg 1 0 50 75 kg 2 1 35 50 kg 3 2 or 1 25 35 kg 4 or 5 3 or 2 15 25 k...

Page 21: ...of the bed sensor under the mattress The correct position is at the chest height of the user and across the bed 3 Power supply battery check if power supply is used Make sure that power supply and bed...

Page 22: ...e device on only after the person has fallen asleep If the bed exit alarm is enabled DIP switch 3 in DOWN position and sensitivity is not set high enough alarm may trigger Either disable alarm DIP swi...

Page 23: ...ntact your dealer distributor or manufacturer To avoid any unnecessary inconvenience on your part we recommend you read these instructions for use carefully before seeking guarantee service 22 1Your g...

Page 24: ...beyond specifications or features described in the instructions for use or modifications to the product to conform it to national or local technical or safety standards in countries other than those...

Page 25: ...xample some countries prohibit the exclusion or limitation of damages resulting from negligence gross negligence wilful misconduct deceit and similar acts Abilia s liability under this guarantee will...

Page 26: ...eset switch 8 pcs DIP switches for settings volume delay one 10 position rotary switch for adjusting sensitivity Signal lights 3 LEDs green blue and red Delays Movement notification delay alternatives...

Page 27: ...W 15 W Output voltage 5 VDC Output current 1 2 A 3 0 A Electrical safety class Class II 23 4Ambient conditions Operating temperature 10 40 C Storage and transport temperature 30 50 C Relative humidity...

Page 28: ...pecified below The customer or the user of the equipment should assure that it is used in such an environment RF emissions CISPR 11 Group 1 The Emfit Tonic Clonic Seizure Monitor uses RF energy only f...

Page 29: ...61000 4 5 1 kV line s to line s IEC 60601 1 2 test level Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Page 30: ...each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may...

Page 31: ...to frequency of transmitter 150 kHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 5 GHz d 1 2 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m...

Page 32: ...32 26 Appendix Related pictures...

Page 33: ...33...

Page 34: ...34...

Page 35: ...35...

Page 36: ...36 464000 464031_Manual Emfit Tonic Clonic Seizure Monitor_EN_RevF1 docx Dec 2017...

Reviews: