31
Recommended separation distances between portable and mobile RF
communications equipment and the Emfit Tonic-Clonic Seizure Monitor.
The Emfit Tonic-Clonic Seizure Monitor is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the Emfit epileptic seizure alarm can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Emfit Tonic-Clonic Seizure Monitor alarm as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of
transmitter
150 kHz
– 80 MHz
d
=
1,2
√
P
80 MHz
– 800 MHz
d
=
1,2
√
P
800 MHz
– 2.5 GHz
d
=
1,2
√
P
0,01
0,12 m
0,12 m
0,23 m
0,1
0,38 m
0,38 m
0,73 m
1
1,2 m
1,2 m
2,3 m
10
3,8 m
3,8 m
7,3 m
100
12 m
12 m
23 m
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
25. Manufacturer
Abilia AB
Box 92
191 22 Sollenuna, Sweden
Phone: +46 8 594 694 00
Email: [email protected]
Internet:
www.abilia.com
Summary of Contents for 464000
Page 2: ...2...
Page 32: ...32 26 Appendix Related pictures...
Page 33: ...33...
Page 34: ...34...
Page 35: ...35...
Page 36: ...36 464000 464031_Manual Emfit Tonic Clonic Seizure Monitor_EN_RevF1 docx Dec 2017...