11
6. Setting up the DIP switches
Open the lid of the control unit by lifting it from the side (picture 9). Inside, you will find
eight DIP switches that are used to select the desired functions (picture 4). Remember to
select the desired settings using the DIP switches before using the device.
The control unit has the following factory settings. DIP 1, 2, 4, 5, 6, 7 and 8 OFF (down).
DIP 3 ON (up). With these settings:
Movement notification delay is 13 seconds.
The SW1 switch acts as the On/Off switch which means that the device can be
switched on or off by pressing the switch for three (3) seconds.
The notification sound is loudest.
The bed exit notification is disabled.
Remove the batteries and disconnect the device from the power supply and
remove nurse call cable (if used) from connector X2 before opening the lid.
Remove the batteries and disconnect the power supply before setting up the DIP
switches in order to implement the new settings.
6.1 Adjusting the tonic-clonic seizure notification delay (switches 1 & 2)
When the device detects faster movement (between 3
–20Hz), its internal timer switches
on. The device gives a notification if the movement continues for longer than the preset
time. The desired delay should be set up according to the table below.
The factory setting is 13 seconds. If the user makes other habitual movements, such as
rocking, scratching or restlessness, a longer delay time can be selected to avoid false
notifications.
Setting the fast movements duration after
which the notification goes off
Switch 1
Switch 2
10 seconds
ON (up)
ON (up)
13 seconds (default)
OFF (down) OFF (down)
16 seconds
ON (up)
OFF (down)
20 seconds
OFF (down) ON (up)