a) Adding a scan profile 

To add a profile, click the 

Add

 button to create a blank profile called “

New Profile

”.  Select the profile in the profile list and 

carry out the following steps : 

 

Enter a suitable 

Profile Name

 – Usually something that describes this profile’s intended use to a user. 

 

Select a suitable 

Base Voltage Range

 from the list, for the device(s) you are going to test. 

 

Set the 

Low Peak Voltage

 – This is the lowest voltage used during the scan, and can be any value which is within 

the 

Base Voltage Range

.

 

 

Set the 

High Peak Voltage

 – This is the highest voltage used during a scan, and can be any value which is within the 

Base Voltage Range

.

 

 

Select a suitable 

Waveform

 type from the list, for the device(s) you are going to test – We recommend using a 

Sine

 

Waveform

, but others are available.

 

 

Select a suitable 

Source Resistance

 from the list, for the device(s) you are going to test.

 

 

Select a suitable 

Frequency

 from the list, for the device(s) you are going to test.

 

The 

Scan Profile

 is now complete, and you can save and exit by clicking the 

OK

 button or continue to editing profiles. 

 

b) Copying a scan profile 

To copy a profile select it from the profile list and click the 

Copy

 button to create a blank profile called “

Copy of

” your selected 

Scan Profile

 name.  Select the profile in the profile list and it is then possible to change the properties of the profile. You can 

save and exit by clicking the 

OK

 button or continue to editing profiles. 

 

c) Editing a scan profile 

To edit a profile select the profile in the profile list and it is then possible to change the properties of the profile.  You can save 
and exit by clicking the 

OK

 button, or you can edit another profile. 

Note only 

User

 profiles can be edited, if you wish to edit a 

System

 profile you must first make a copy of it.  

 

d) Deleting a scan profile 

To delete a profile select the profile in the profile list and click the Delete button. This will remove the profile from the list. A 
scan profile cannot be deleted if at least one device in the library is using it. 

Note that the profile is not permanently removed until the 

OK

 button is clicked. 

Summary of Contents for SENTRY

Page 1: ...tector OPERATOR S MANUAL ABI Electronics Ltd Dodworth Business Park Barnsley S75 3SP United Kingdom 44 0 1226 207420 44 0 1226 207620 sales abielectronics co uk www abielectronics co uk Copyright 2012 2017 ABI Electronics Ltd ...

Page 2: ...nnection 6 b Input ratings 6 c SENTRY software installation 6 d USB driver installation 6 3 3 Running the software 6 3 4 Quick start guide 7 a Learning a device 7 b Testing a device 7 c Combining PinPrints 7 IV LIBRARY MANAGEMENT a Adding a device to the library 9 b Editing a device 9 c Viewing a device 9 d Deleting a device 9 e Exporting a device 9 f Importing a device 10 g Importing exporting th...

Page 3: ...Adding a scan profile 21 b Copying a scan profile 21 c Editing a scan profile 21 d Deleting a scan profile 21 VIII OTHER FUNCTIONS 8 1 Choosing reference mode 22 8 2 Adding device documentation 22 8 3 Configuration 23 a User administration 23 b Hardware administration 23 8 4 Calibration 24 8 5 Compensation 24 8 6 Comparison tolerances 25 a Comparison point tolerance 25 b Pin comparison tolerance 2...

Page 4: ...uipment is used in conjunction with other equipment not supplied by ABI We recommend that you use ABI supplied accessories where possible The SENTRY is designed to operate in Residential Commercial and Light Industrial environments and is not approved for operation in heavy industrial medical automotive or aircraft environments unless confirmed in writing by ABI c Location The SENTRY should be ins...

Page 5: ...marked to resemble more expensive ICs ICs salvaged from circuit boards It is getting more and more difficult to ensure that the components you purchase are genuine The vast majority of dealers world wide are honest and genuine but there are unfortunately some rogue dealers who will try to make money by attempting to sell non genuine components c SENTRY hardware operation The ABI SENTRY is an elect...

Page 6: ...ects to a standard PC via a high speed USB interface The SENTRY operating software on your PC controls all system operations and functions and provides full database facilities for storage and reporting of PinPrints from genuine devices and manufacturers e PC requirements The ABI SENTRY control software runs on a PC with the following minimum specifications Microsoft Windows XP Vista or 7 operatin...

Page 7: ...een You may see a security warning during the installation d USB driver installation Connect the SENTRY unit to a suitable mains supply and switch on then connect the unit to a spare high speed USB 2 0 port on your PC with the cable supplied After a few seconds you will see the message Found New Hardware ABI SENTRY Counterfeit IC Detector You may also see a warning message if the USB port selected...

Page 8: ...indication of date time and user will be displayed in the PinPrint section Note that the Clear button can be used to delete saved PinPrints The acquisition of reference PinPrints is now complete For the purpose of this guide leave the IC in the socket and move to the next section by clicking Test from the left hand side menu b Testing a device This section is aimed at testing devices by comparing ...

Page 9: ... devices which match your criteria This can also be used in conjunction with the device type filter to further refine a search for a specific device The device type filter By selecting a type from the list the SENTRY software will search through your device library and return only devices which match this type This can also be used in conjunction with the device text filter to further refine a sea...

Page 10: ...can accurately scan the device You have now completed adding a new device Click OK to save the device into your library b Editing a device Note the Edit button is only visible if the device is eligible for editing If you want to change any of the settings for a device select the device from the list and click the Edit button You can then change any of the settings as described in the section above...

Page 11: ... library click on the Select All button in the library section and click on Export The devices will be exported individually to a destination of your choice To import multiple devices to a library click on Import in the library section and browse to the location of the devices on your computer Hold the CONTROL key down on the keyboard and select the devices required Alternatively use the CONTROL S...

Page 12: ...his can also be used in conjunction with the other filters to further refine a search for a specific package The description filter By typing part or all of a package description the SENTRY software will search through your package library and return only packages which match that description This can also be used in conjunction with the other filters to further refine a search for a specific pack...

Page 13: ...pe as the previous pin added The properties of the pin size location shape number are displayed in the property grid at the bottom of the window when a pin is selected You can change the properties of the pin by resizing with the mouse and or by using the controls in the pin property grid Add shape tool for adding shapes to your package The properties of the shape size location colour style are di...

Page 14: ...ight click on the pin which you want to be the next number in the sequence The number of the pin will be set to the next number down from the previous one Continue double right clicking on all pins in the order you want to renumber the entire package Automatic renumber This function will start from the pin you select and renumber all higher numbered pins in sequence This means you can quickly crea...

Page 15: ...the extension PKG will be created in the chosen location To export multiple packages hold the CONTROL key down on the keyboard and select the packages to be exported Alternatively use the CONTROL SHIFT keys to select a complete list of packages Click Export to export the packages individually h Importing a package To import a package click the Import button and browse to the PKG file for the packa...

Page 16: ...n conjunction with the other filters to further refine a search for a specific Adapter The package type filter By selecting a type from the list the SENTRY software will search through your Adapter library and return only those which have packages which match this package type This can also be used in conjunction with the other filters to further refine a search for a specific Adapter The descript...

Page 17: ...you insert a 14 pin DIL IC in the rear of the socket this means that pin 1 of the IC connects to channel 1 of the SENTRY pin 2 to channel 2 and so on However pin 14 of the IC on the other side connects to channel 48 of the SENTRY as shown in the table Pin 13 connects to channel 47 and so on There is an Adapter included as standard for this scenario called 14 pin DIL in 48 pin ZIF This Adapter is u...

Page 18: ...guidance when inserting ICs The label will resize itself automatically depending on the text and chosen font You can change the location of the label by dragging with the mouse and or by using the controls in the label property grid Add package tool for adding pre defined packages to your Adapter The location of the package is displayed in the property grid at the bottom of the window when a packa...

Page 19: ...m the previous one Continue double right clicking on all pins on all packages in the order you want to allocate channel numbers for the entire Adapter c Editing an adapter Note the Edit button is only visible if the Adapter is eligible for editing If you want to change any of the information or design for an Adapter select it from the list and click the Edit button You can then change the informat...

Page 20: ...dapters to be exported Alternatively use the CONTROL SHIFT keys to select a complete list of adapters Click Export to export the adapters individually h Adapter files The packages you design are stored as text files in the Adapters folder on your computer hard drive These files are human readable but you should not attempt to edit them as you may cause unexpected problems and your changes may be o...

Page 21: ...lected profile to the library see section 0 for more information Note this button is only enabled if a profile is selected in the profile list Error Reference source not found The Delete button This button allows you to remove a profile from the library Note this button is only enabled if a profile is selected in the profile list The Hardware status Shows the current status of the connected hardwa...

Page 22: ...or the device s you are going to test The Scan Profile is now complete and you can save and exit by clicking the OK button or continue to editing profiles b Copying a scan profile To copy a profile select it from the profile list and click the Copy button to create a blank profile called Copy of your selected Scan Profile name Select the profile in the profile list and it is then possible to chang...

Page 23: ... pin count devices these modes take more time 8 2 Adding device documentation In the Learn section click on the Documentation tab In this section you can add various documents and graphics to help in identifying counterfeit ICs You can add pictures either from a file or from a USB camera connected to your PC and documents such as device data sheets etc You can add any type of document that you can...

Page 24: ...apters Manage Scan Profiles this permission allows the user to change the hardware settings used when performing scans Adding a user To add a new user click on the Administration button then click Add Enter a name for the new user and set the access permissions you require by checking or unchecking them in the permissions list If you wish you can set a password for the new user by clicking the Set...

Page 25: ...y requires the user to enter a few voltage readings in the initial stages Continue to follow the on screen instructions until calibration is complete When calibration is completed click Next to save the calibration to the SENTRY unit at this point the SENTRY is automatically reset and should be ready for use immediately If you experience problems after calibration power down the SENTRY wait 5 seco...

Page 26: ...al tolerance The percentage of the total range that the measurement needs to be within in the horizontal direction for the data to be deemed valid Vertical The percentage of the total range that the measurement needs to be within in the vertical direction for the data to be deemed valid It is possible to restore the default settings for these by clicking the Restore Default button which sets all t...

Page 27: ...ors can also be used Component Reference images If images and photos have been added to the device documentation see section 8 2 they can also be added to the report Select the image from the list available and with the left mouse key still pressed drag the image to the area under component or reference as required Resize functions All images diagrams and PinPrints can be resized as required The R...

Reviews: