EDUCATIONAL MATERIAL FOR RISK MINIMISATION
Display
Alarm
Cause
Action
Upstream
Occlusion
Two-tone
alarm signal
Fluid is not flowing
from the cassette to
the pump�
Check for a kink
in the tubing or
a closed clamp
between the fluid
container and pump�
Press NEXT or STOP/
START to stop the
pump and silence
the alarm for two
minutes� Remove
the obstruction and
press STOP/START
to restart the pump�
If necessary, remove
the cassette from the
pump and connect
it again�
No
Disposable,
Clamp
Tubing
Two-tone
alarm signal
The cassette has
been removed�
Close the tube
clamp immediately
or disconnect it from
the tube� A cassette
must be connected
before the pump can
be started
Press STOP/START
or NEXT to stop the
alarm� Very cold or
viscous medicine has
caused the alarm�
Allow medicine
to reach room
temperature before
connecting it to the
pump� Check if the
cassette is empty –
change it if necessary�
Power Lost
While Pump
was On
Two-tone
alarm signal
The pump was
running when power
was removed�
Stop the pump before
changing the batteries
or removing the power
source� Press NEXT or
STOP/START to stop
the pump and turn off
the alarm�