EDUCATIONAL MATERIAL FOR RISK MINIMISATION
Display
Alarm
Cause
Action
Battery
depleted
Two tone alarm
The battery power
is too low to
operate the pump�
The pump is now
stopped�
Change the
batteries
immediately� Press
and hold STOP/
START to restart the
pump�
Battery
Removed
Two tone alarm
The batteries have
been removed
while the pump
is running, or you
have tried to start
the pump with
depleted batteries�
The pump is now
stopped
Press STOP/START
or NEXT to stop
the alarm� Reinstall
batteries or install
new batteries
Press and hold
STOP/START to
restart the pump�
Remote
Dose Cord
Removed
Two single-tone
signals if the
pump is stopped�
Two-tone alarm
signal if the
pump is on�
The remote dose
cord was removed
Reinsert connector
or press NEXT to
silence the alarm�