This document is a guide. For complete details, always refer to the NT2000iX
™
RF Generator Operator’s Manual.
EDITING TREATMENT SETTINGS
FROM THE WELCOME SCREEN
1. Tap
Doctor.
2. On the left, tap the treatment mode you want to
edit. The settings for that treatment then appear on
the right. To select a different treatment mode to
edit, first tap
Save
or
Cancel.
FROM THE LESION PROCEDURE SCREEN
1. Tap the treatment settings button in the lower
left corner to change the settings for the current
procedure only.
DEFAULT SETTINGS
•
Thermal Lesion RF default: 80 °C for 60 seconds
•
Pulsed RF default: 42 °C for 2 minutes with a pulse
width of 20 ms and a frequency of 2 Hz
•
The Default doctor profile settings cannot be
edited. To edit preferences, a new profile must
be created.