29
RF telemetry operates:
Using the MICS band (402-405 MHz)
At 200 kilobits/s or lower
Over a short range (less than 2.5 meters in normal use)
Bluetooth
®
Low Energy wireless communication operates:
At 2.402 to 2.4835 GHz at 1 megabits/s or lower
Over a short range (within 2 meters in normal use)
During use, the Merlin 2 PCS inductive telemetry wand displays a signal strength indicator to help
clinicians optimize telemetry wand placement. The transmitter automatically monitors and
recovers from most communication errors, including data integrity errors and excessive latency. If
wireless Quality of Service (QoS) is inadequate, the programmer automatically switches to
inductive telemetry.
Transportation and Handling Conditions
Table 7. Transportation and handling conditions
Minimum Temperature (°C (°F))*
-25 (-13)
Maximum Temperature (°C (°F))*
70 (158)
Minimum Humidity (% non-condensing)*
10
Maximum Humidity (% non-condensing)*
90
Minimum Atmospheric Pressure (hPa)
500
Maximum Atmospheric Pressure (hPa)
1060
*Found on box label
Operation and Storage Conditions
Table 8. Operation and storage conditions
Minimum Temperature (°C (°F))*
10 (50)
Maximum Temperature (°C (°F))*
35 (95)
Minimum Humidity (% non-condensing)*
30
Maximum Humidity (% non-condensing)*
75
Minimum Atmospheric Pressure (hPa)
700
Maximum Atmospheric Pressure (hPa)
1060
*Found on Merlin™ 2 PCS label
Disposal
Return the Merlin™ 2 PCS and the Merlin Antenna to Abbott Medical at the end of their operating
lives.
The crossed-out trash can symbol marked on the Merlin 2 PCS and the Merlin Antenna indicates
that Abbott Medical complies with the European Union’s Waste Electrical and Electronic