1000SR System Quick Troubleshooting Guide

Error Code 5209  No Reaction Vessel (RV)
Detected in Process path after RV Load 
Attempt

• If RV loader chute is not in position to load RVs, 

reseat the RV loader chute into the wash zone 
manifold.

Chute is in the 

UP

position (incorrect).

Chute is in the 

DOWN

position (correct).

PN 205854-101

©

2009, Abbott Laboratories, Abbott Park, IL, 60064

Refer to the Operations Manual for further instructions

Level Sense Errors:  3000

• Ensure sample cup or tube is present and adequate 

sample is in the cup or tube.

Reagent probe areas only:

• Reagent bottle is empty:  Replace empty reagents.
• WZ probe maintenance water bottle is missing, over-

filled or filled with distilled (DI) water instead of tap 
water or saline or bleach:  Repeat the maintenance 
procedure, ensuring the WZ probe maintenance water 
bottle is in position and contains the correct volume 
and solution.

Sample Pipettor:

Tighten the probe clamp finger tight.

Only fill maintenance
bottle 25 to 30ml full.
Errors may occur if
filled with more than
30ml of solution.  

 

®