1000SR System Quick Troubleshooting Guide

Error Code 3700 Unable to Process Test, (x)
Wash Aspiration Error for Probe(s) (y)

PN 205854-101

©

2009, Abbott Laboratories, Abbott Park, IL, 60064

Refer to the Operations Manual for further instructions

Customer Support:  1-877- 4ABBOTT  (1-877- 422-2688

)

Instrument SN:  ______________

• Remove fibrin clots from sample cup or tube with a 

clean applicator stick.

• Perform the appropriate 

as-needed

maintenance 

procedure:

9

6445 Pipettor/WZ Probe Cleaning

• Replace the wash zone probe.
• Replace the wash zone temperature tubing and 

sensor.

Wash Zone Tubing and Probe Assembly

Touchscreen Monitor fails to respond

Cycle power to System Control Center (SCC) and to the 

Processing module by performing the following:

• Select 

F3-Shutdown

on the Snapshot screen.  A 

confirmation message displays.

• Select 

OK

to confirm the shutdown.

• Wait for the information window to display, and then 

simultaneously press the 

CTRL+ALT+DELETE

keys 

on the keyboard.  The Confirm Exit window displays.

• Leave the 

Shutdown Computer

option selected, 

select 

OK

, and then wait for the information window 

to display.

Turn off the power to the processing module power 
supply by moving the power switch down.

NOTE:

The Module Power 

Switch is located on the
back side of the Processing
Module.

If touch-screen connector is loose or disconnected, 
reseat or reconnect the touch-screen connector.

[1] monitor power cable

[2] touch-screen cable
[3] monitor video cable

• Press the power switch on the front of the CPU to 

turn on the SCC.

• Wait for the Snapshot screen to display.  It may 

take several minutes to display.

• Ensure the processing module(s) has been 

powered off for one minute, then move the module 
power supply power switch up to turn on power.

1 2 3

Page 1 of 2

Locate the Central Processing
Unit (CPU).

NOTE:

Open the card cage and 

SCC center door to access the 
power switch.

Press and hold the power switch
on the front of the CPU to turn off
power to the SCC.

NOTE:

The SCC may power off 

immediately, or it may take up to 
10 seconds depending on the 
type of SCC you have.