CAUTION: During the installation, do not remove the instrument or
the USB memory device from the Base Station. When the installation
is complete, the instrument will display a message indicating that new
software has been installed. The display of this message is the indication
that it is safe to remove the instrument and USB memory device from the
Base Station.
52
i-STAT
Alinity — Quick Reference Guide
Art: 731848-01 Rev. M Rev. Date: 25-Jul-2022
SECTION 10
HOW TO PERFORM SOFTWARE UPDATE (CONT.)
INSTALLATION
1
2
Ensure that the prerequisites, on the previous page have been completed.
Place instrument into the
base station and power-up
instrument.
Insert USB into the USB
port in the base station.
3
From the Home Screen:
> More Options
> Instrument Options
> Software Installation
> Install from USB
5
OSi Software
Installation Results:
The instrument automatically restarts upon successful
installation!
4
Follow prompts until
transfer is complete.
Refer to onscreen details for
cause and repeat procedure.
If failure persists,
contact your local Abbott
representative.