![background image](http://html2.mh-extra.com/html/abbott/i-stat-alinity/i-stat-alinity_quick-reference-manual_3351098038.webp)
SECTION 6
HOW TO PERFORM QUALITY TESTING (CONT.)
LIQUID QUALITY CONTROL (CONT.)
34
i-STAT
Alinity — Quick Reference Guide
Art: 731848-01 Rev. M Rev. Date: 25-Jul-2022
Once customized, the results will appear as follows:
Numeric value with no arrow – In Range
Numeric value with arrow – Out of Range High
Numeric value with arrow – Out of Range Low
BEST PRACTICE EXAMPLE
Use AlinIQ CWi to customize for automatic pass/fail
determination using eVAS.
AnGap, mmol/L
34
Op�ons
Menu
Page
Na, mmol/L
129
127 - 136
Home
View
Entered Info
Op�ons
Menu
Transmit
i-STAT CHEM8+
i-STAT TriControl L2: 321068
PASS
16FEB2025 11:01
BUN, mg/dL
10
8 - 14
Crea, mg/dL
1.3
0.7 - 1.5
Glu, mg/dL
91
85 - 100
K, mmol/L
3.8
3.5 - 4.1
iCa, mmol/L
1.33
1.24 - 1.42
Cl, mmol/L
92
85 - 98
TCO2, mmol/L
18
16 - 29
Op�ons
Menu
Page
i-STAT CHEM8+
i-STAT TriControl L1: 321054
FAIL
16FEB2025 09:45
Na, mmol/L
124
117 - 126
BUN, mg/dL
66
47 - 62
Glu, mg/dL
280
228- 312
K, mmol/L
3.0
2.6 - 3.2
Crea, mg/dL
3.7
3.0 - 4.6
iCa, mmol/L
0.80
0.73 - 0.89
AnGap, mmol/L
42
Cl, mmol/L
72
68 - 77
TCO2, mmol/L
10
8 - 26
Home
View
Entered Info
Op�ons
Menu
Transmit