14-10
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Retic Menu Options
Section 14
Turning OFF the Reticulocyte Package
The
[TURN OFF RETIC PKG]
key is located on the
RETIC
OPERATION SET UP MENU
screen (see the Figure 14.2), which is accessed from the
RETIC SET UP
screen. The
Reticulocyte Package software is disabled when the
[TURN OFF RETIC PKG]
key is
pressed. When the Reticulocyte Package is OFF, the soft keys are displayed as
described in
Chapter 5:
Operating Instructions
.
Figure 14.2 Operation Set Up Menu Screen with Reticulocyte Package Enabled
When the Reticulocyte Package is turned OFF, the instrument automatically runs a
wash cycle before returning to the Standard Hematology Mode.
CAUTION:
If the instrument has been idle for four hours, it enters the
STANDBY
state and automatically returns to the Standard Hematology mode
without
running a cleaning cycle. If this happens, perform an Auto-Clean
cycle before using the instrument again.