10-36
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
No Entry Found
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The number (sequence number or specimen ID number) entered on the
DATA
LOG SEARCH
screen is not present in the data log.
Check that the entry was correct. If appropriate, enter the correct number.
Duplicated Specimen ID on the new line
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The Work List already contains a bar code number that has been entered.
It is not possible to enter the same bar code number twice in the Work List. If
appropriate, delete the previous entry and re-enter the number.
Sample Loader is busy
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
An action was requested during Sample Loader operation and the Sample Loader
cannot perform it.
Press the
[STOP LOADER]
soft key before requesting the desired action.
Change Sampler in Ready State Only
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The
[CHANGE SAMPLER]
key was pressed while the Analyzer was busy.
The
[CHANGE SAMPLER]
key can only be pressed when the Analyzer is in the
READY state.
Change Sampler when Sample Loader is not busy
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The
[CHANGE SAMPLER]
key was pressed while the Sampler Loader was busy.
Press the
[STOP LOADER]
soft key before pressing
[CHANGE SAMPLER]
.
Loader Status 143: samples completed
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The last rack in the load zone has been processed. NOTE: The Sample Loader
halts.
No corrective action is necessary. When you wish to run additional samples in
the Sample Loader, load racks and press the
[START LOADER]
key.
Table 10.1 Status
Conditions (Continued)