CELL-DYN
®
3200 System Operator’s Manual
10-15
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Figure 10.14 Drain Accumulators Screen
Drain Accumulator
When the
[DRAIN ACCUMULATOR]
key is pressed, the internal vacuum
accumulators are drained of accumulated fluid. This key is used to correct the
“Vacuum Accumulator Wet” fault. The screen displays (see Figure 10.14) the
following message:
After draining the accumulators, press INITIALIZATION
key to return to normal operation.
When the process is completed, the system must be initialized and primed. See
Initialization
below.
Initialization
When the
[INITIALIZATION]
key is pressed, the Analyzer is initialized. This is
necessary when a fatal fault has occurred.
After the Analyzer is initialized, it must be primed. To prime the Analyzer, go to
the
MAIN MENU
and press the
[RUN]
soft key.