Performance Characteristics and Specifications
Performance Characteristics
Section 4
Typical Paired Difference Precision Normal
The typical paired difference precision of the measured parameters is presented as
the total CV calculated from samples run in duplicate on the same system. (Refer
to Table 4.15.)
Sensitivity and Specificity of WBC Differential Flags
The sensitivity and specificity of the WBC flags were evaluated using the general
principles for morphological classifications outlined in the NCCLS H20-A
Standard: “Reference Leukocyte Differential Count (Proportional) and Evaluation
of Instrumental Methods” as a guideline. The statistics were determined by
comparing the CELL-DYN 3200 System results with results obtained from a
reference CELL-DYN 3500 System. Discrepant results were arbitrated with a
manual, microscopic, 400-cell differential count.
Table 4.15 Typical Precision of Hemogram Parameters
Parameter
Number of Pairs
Total CV
WBC (WOC)
10
2.01%
RBC
10
0.81%
HGB
10
0.72%
MCV
10
0.38%
RDW
10
1.76%
PLT
10
3.11%
MPV
10
3.68%
%N
10
1.23%
%L
10
2.07%
%M
10
7.72%
%E
10
8.06%
%B
10
16.82%