CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
TROUBLESHOOTING
IP - 11 Alarm 4 (e and j) Time-out Moving X-Axis, 2 of 8
1 - 36
1
IP - 11 Alarm 4 (e and j) Time-out Moving X-Axis, 2 of 8
SL-129C
1
2
20
19
TP
[NON- CE MARK]
Pin 7 X-Axis Limit
Pin 3 X-Axis Home
Pin 3
X-Axis Home
CN4
20
19
2
1
[CE MARK]
Pin 15
X-Axis Limit
X-Axis
Home Sensor & Limit
Sensor = +5.0V Blocked,
0.0V Unblocked
(See Figure)
?
Problem
Corrected
?
Perform Total Call
Yes
Replace Defective Sensor
(
No
- Check Sensor
Connections
- Check CPU Main Bd.
CN4 for Proper
Connection
- Ensure Sensor Flag
Passes through Center
of Sensor
Yes