VERIFICATION PROCEDURESVP - 7.1 Mixing Paddle Overcurrent Alarm Check
3 - 115
3
6. Press R on keyboard to turn on mixing assembly.
7. Verify Loader beeps.
• For F or G Loader go to Step 11.
• For
E
Loader
go to Step 9.
8. If check fails, go to Step 10.
9. Perform the following procedure (E Loader only):
a. Reconnect Motor # 1 and connect Mixing Motor Current Tool to Mixing Motor #2.
b. Press (R) key.
-
If Loader beeps, go to Step 11.
-
If check fails, go to Step 10.
10.Perform
“VP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment” (Page 3-116)
.
11.Connect motor.
E Loader Only - Check Connection Match up.
a. Mixing Motor # 1 should be connected to Blue and Purple wires.
b. Mixing Motor # 2 should be connected to Yellow and Orange wires.
c. Set BAR Switch to 0.
d. Press and hold (R) key. Only Mixing Motor #1 (left) rotates.
--- END ---