28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment 3 - 116
3
VP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment
Purpose: Perform this procedure to adjust Mixing Paddle Overcurrent Alarm.
NOTE:[EUROPE] Perform this adjustment on CE MARK and NON-CE MARK loaders
[REST OF WORLD] Perform this adjustment on NON-CE MARK Loaders only.
1. Perform Check procedure
“VP - 7.1 Mixing Paddle Overcurrent Alarm Check” (Page 3-113)
.
2. Set up Mixing Paddle (
“VP - 7.1 Mixing Paddle Overcurrent Alarm Check” (Page 3-113)
1-6.
3. Adjust Overcurrent Alarm as follows:
a. [CE MARK] Europe Only Ensure TP 15 is cut (
b. Adjust Potentiometer fully counter clockwise (
[CE MARK] Europe Only - R64
[NON-CE MARK] - VR10
c. While pressing “R” key, adjust Potentiometer clockwise until Loader just begins to beep.
d. If F Loader, disconnect Resistor Tool and reconnect Motor. Procedure is complete. If E Loader
Go to Step e.
Special Tools
DESCRIPTION
QTY
PN
Mixing Motor
Current Tool
1
9450121