CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 7.1 Mixing Paddle Overcurrent Alarm Check
3 - 113
3
VP - 7.1 Mixing Paddle Overcurrent Alarm Check
Purpose: Perform this procedure to verify Mixing Paddle Overcurrent Alarm.
NOTE:[NON-CE MARK] only. [CE MARK] is not adjustable.
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Set BAR Switch to 1.
3. Remove Tower Cover.
4. Disconnect Mixing Motor In-Line connector. If no In-Line connector, unsolder at Mixing Motor (E
Loader - Disconnect Motor # 1(Left),
Special Tools
DESCRIPTION
QTY
PN
Mixing Motor
Current Tool
1
9450121