CELL-DYN
3000 System Operator’s Manual
9-33
9140240E — May 1995
Chapter 9
Maintenance
Figure 9.12:
Special Protocols: Extended Auto-Clean Screen
Procedure
1.
Ensure that waste container is less than 1/2 full.
2.
Select the open sampler mode.
3.
Carefully wipe the outside of the open sample aspiration probe and
the bottom of the wash block with a piece of gauze dampened with
warm water and a few drops of CELL-DYN Enzymatic Cleaner.
Wipe any dried reagent or blood off the bottom of the wash block.
4.
Press [SPECIAL PROTOCOLS] followed by [MORE] to access
the Extended Auto-Clean function.
5.
Press [EXTEND AUTOCLEAN].
Instructions for performing the procedure are given on the screen
(
6.
Dispense approximately 1.5 mL of enzymatic cleaner undiluted
into a clean container and hold the container under the open sample
aspiration probe.
NOTE: Do not press the touch plate. The Extended Auto Clean
procedure is initiated only by the [START] key.
START
CANCEL
SPECIAL PROTOCOLS
Ready
Nov 091992
Operator ID
Sequence #
12:25
734
0630
Hold a bottle of Cell-Dyn Enzymatic Cleaner under the probe,
and press the START key
When the Extended Auto-Clean process is completed,
the instrument will automatically go to STANDBY.
This process takes 2.5 hours.
NOTE:
After 38 minutes, the cleaning process may be interrupted
to process specimens.