19
4.7C Carbon Monoxide Monitor Input – Fig. 4.3C (overleaf)
Connect the milliamp source in place of the 'CO I/P' connections. Switch on.
Fig. 4.2C Analogue Retransmission, Electrical Calibration
%CO
2
(Inferred)
CO
2
/
CO
%CO
2
not available
Select the Carbon Dioxide/Carbon Monoxide
page.
or
4 Times
Advance to electrical calibration procedures.
▲
Operate for approximately three seconds to gain access to the electrical
calibration procedures within the Carbon Dioxide/Monoxide page.
N o t e .
'x x x x x x' in the following illustrations is a numerical code used for
software analysis during manufacture and can be disregarded.
Continued overleaf.
Refer to Overall Programme
Chart – Fig. 10 in ZMT/0011
ANALOGUE RETRANSMISS
With Security Access
Without
Security
Access
OP1 ReTx %O
2
off
on
Cell T
Flue T
Air T
Ef 'ncy
OP1 ReTx %O
2
or
Cell T
Flue T
Air T
Ef 'ncy
➞
Vary Lo OP1 & ENT
➞
➞
Vary Hi OP1 & ENT
➞
➞
➞
➞
off or
OP1 ReTx on or
SECURE PARAMETERS
Omitted if respective
thermocouple
temperature is
preset
OP1 Zero x x x x x x
OP1 Span x x x x x x
Analogue Retransmission
Calibration Parameters
OP2 ReTx %O
2
off
on
Cell T
Flue T
Air T
Ef 'ncy
Omitted if respective
thermocouple
temperature is
preset
Single retransmission versions
Repeat for other
retransmission
output(s), as
applicable