28
VA Master FAM540
METAL CONE VARIABLE AREA FLOWMETER | OI/FAM540-EN REV. E
… 3 Use in potentially explosive atmospheres in accordance with FM and cCSAus
… Safety specifications FM, cCSAus
Table 2: Analog indicator with alarm signaling unit (temperature data for FM in °F, for cCSAus in °C)
Order
code*
Marking Terminals
Electrical
values
T
amb
−
58 °F to
T
amb
−
50 °C to
Temp. class
T
medium
Maximum
Insulation
F3 FM
XP / CL I / DIV 1 / GP ABCD / T6 ... T1
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1
CL I, ZN 1 AEx d IIC T6 ... T1
cCSAus
XP / CL I / DIV 1 / GP BCD / T6 ... T1
DIP / CL II, III / DIV 1 / GP EFG / T6 ... T1
Ex d IIC T6 ... T1
DIP A21 TA 85°C ... T
medium
41 / 42
**
51 / 52
**
U
max
= 16 V
I
max
= 25 mA
P
max
= 64 mW
104 °F
40 °C
T1
824 °F
440 °C
No
104 °F
40 °C
T1
707 °F
375 °C
yes
104 °F
40 °C
T1
500 °F
260 °C
yes
122 °F
50 °C
T1
572 °F
300 °C
yes
122 °F
50 °C
T2
554 °F
290 °C
yes
122 °F
50 °C
T2
428 °F
220°C
yes
140 °F
60 °C
T2
608 °F
320 °C
No
140 °F
60 °C
T2
446 °F
230 °C
yes
140 °F
60 °C
T3
338 °F
170 °C
yes
158 °F
70 °C
T3
383 °F
195 °C
No
158 °F
70 °C
T3
302 °F
150 °C
yes
yes
yes
yes
158 °F
70 °C
T4
266 °F
130 °C
158 °F
70 °C
T5
203 °F
95 °C
140 °F
60 °C
T6
176 °F
80 °C
41 / 42
**
51 / 52
**
U
max
= 16 V
I
max
= 52 mA
P
max
= 169 mW
104 °F
40 °C
T1
824 °F
440 °C
No
yes
yes
yes
yes
yes
No
104 °F
40 °C
T1
707 °F
375 °C
104 °F
40 °C
T1
500 °F
260 °C
122 °F
50 °C
T1
572 °F
300 °C
122 °F
50 °C
T2
554 °F
290 °C
122 °F
50 °C
T2
428°F
220°C
140 °F
60 °C
T2
608 °F
320 °C
140 °F
60 °C
T2
446 °F
230 °C
yes
140 °F
60 °C
T3
338 °F
170 °C
yes
158 °F
70 °C
T3
383 °F
195 °C
No
158 °F
70 °C
T3
302 °F
150 °C
yes
158 °F
70 °C
T4
266 °F
130 °C
yes
140 °F
60 °C
T5
140 °F
60 °C
yes
122 °F
50 °C
T5
194 °F
90 °C
No
104 °F
40 °C
T6
140 °F
60 °C
yes
41 / 42
**
51 / 52
**
U
max
= 16 V
I
max
= 76 mA
P
max
= 242 mW
104 °F
40 °C
T1
824 °F
440°C
No
104 °F
40 °C
T1
590 °F
310 °C
yes
104 °F
40 °C
T2
374 °F
190 °C
yes
122 °F
50 °C
T2
644 °F
340 °C
No
122 °F
50 °C
T2
446 °F
230 °C
yes
140 °F
60 °C
T2
446°F
230 °C
No
140 °F
60 °C
T3
320 °F
160 °C
yes
158 °F
70 °C
T4
248 °F
120 °C
No
158 °F
70 °C
T4
212 °F
100 °C
yes
104 °F
40 °C
T5
140 °F
60 °C
yes
86 °F
30 °C
T6
86 °F
30 °C
yes
* Order code ‘Explosion protection and approvals’ (versions 9, 10); refer to
Ordering information in the data sheet
on page 62.
** for connection to a non-intrinsically safe circuit