PURPOSE/SCOPE
PR4 - 1
PROCEDURE PR4 - TWO-INCH SANITARY TRI-CLAMP
ASSEMBLY
PURPOSE/SCOPE
This procedure explains how to install the two-inch sanitary tri-clamp
assembly
Parts
Tools
None.
SAFETY CONSIDERATIONS
PROCEDURE
The Type TB4043 Toroidal Conductivity Sensor provides installation capa-
bility for sanitary pipelines and tanks with two-inch tri-clamp fittings.
R
1. Install the sanitary gasket onto the two-inch sanitary tri-clamp tee (Fig.
).
R
2. Slide the sensor into the two-inch sanitary tri-clamp tee.
R
3. Position the sensor so that the sensing area is perpendicular to the flow
path to meet the installation guidelines described in
20 min.
Number
Qty
Description
customer-supplied
1
Gasket, sanitary
customer-supplied
1
Tri-clamp tee, sanitary, 2 in.
customer-supplied
1
Tri-clamp, sanitary
customer-supplied
1
Conduit/cable gland adapter, ½ in.
customer-supplied
A/R
Conduit, flexible, ½ in., or cable gland
WARNING
1. Never exceed the sensor or sanitary clamp temperature and
pressure limits. Exceeding these limits can damage the equip-
ment in such a way as to pose the risk of injury to personnel.
W1
A/R = as required