FMT230, FMT250
THERMAL MASS FLOWMETER | CI/FMT230/250-EN REV. B
2
Use in potentially explosive atmospheres in accordance with ATEX and IECEx
Note
Further information on the approval of devices for use in potentially explosive atmospheres can be found in the type examination
certificates or the relevant certificates at
www.abb.com/flow
.
Device overview
Standard / No explosion protection
Zones 2, 22
Zone 1, 21 (Zone 0)
Model number
FMT2xx Y0
FMT2xx A2
FMT2xx A1, A3
• Standard
• Zone 2, 22
• Zone 1, 21
• Zone 0
Ex marking
Note
• Depending on the design, a specific marking in accordance
with ATEX or IECEx applies.
• ABB reserves the right to modify the Ex-marking. Refer to the
name plate for the exact marking.
Model FMT2xx-A2… in Zone 2, 22
ATEX / IECEx
Certificate (Atex)
FM19ATEX0178X
Certificate (IECEx)
IECEx FMG 19.0025X
II 3G Ex ec mc IIC T6…T2 Gc
II 3D Ex tc IIIC T85°C…T
medium
Dc
Model FMT2xx-A1… in Zone 1, 21
ATEX / IECEx
Certificate (Atex)
FM19ATEX0177X
Certificate (IECEx)
IECEx FMG 19.0025X
II 2G Ex eb ia mb IIC T6...T2 Gb
II 2G Ex ia IIC T6...T2 Gb
II 2D Ex ia tb IIIC T85°C...T
medium
Db
permitted supply short-circuit current: 35A
Model FMT2xx-A3… in Zone 0, 1, 21
ATEX / IECEx
Certificate (Atex)
FM19ATEX0177X
Certificate (IECEx)
IECEx FMG 19.0025X
II 1/2 G Ex eb ia mb IIC T6...T2 Ga/Gb
II 2G Ex ia IIC T6...T1 Ga
II 2D Ex ia tb IIIC T85°C...T
medium
Db
permitted supply short-circuit current: 35A
Temperature data
Temperature resistance for the connecting cable
The temperature at the cable entries of the device is dependent
on the measuring medium temperature T
medium
and the ambient
temperature T
amb.
.
For the electrical connection of the device, use only cables with
sufficient temperature resistance in accordance with the
following table.
T
amb.
Temperature resistance for the connecting cable
≤
50
°C (≤
122 °F)
≥
70
°C (≥
158 °F)
≤
60
°C (≤
140 °F)
≥
80 °C
(≥
176 °F)
≤
70 °C
(≤
158 °F)
≥
90 °C
(≥
194 °F)
From an ambient temperature of T
amb.
≥
60
°C (≥
140 °F), the
wires in the connection boxes with the enclosed silicone hoses
need to be additionally insulated.
Note
The signal cable supplied by ABB can be used without
restrictions up to an ambient temperature of ≤
80
°C (≤
176 °F).