Commissioning
16 Sensyflow
FMT700-P
42/14-42-EN
Pos: 10 /Überschriften/1/G - I/Inbetriebnahme @ 0\mod_1131980668835_3101.doc @ 3150 @ 11
4 Commissioning
Pos: 11 /Inbetriebnahme/Durchfluss/FMT700-P/Inbetriebnahme @ 33\mod_1253104112968_3101.doc @ 301157 @
1.
Fit the flowmeter sensor in the measuring section. Ensure that the tapered flange connection
is secure (tightness) and observe the flow direction (direction indicated by arrow).
2.
Connect the flowmeter sensor and supply / evaluation unit using the flowmeter sensor cable
(7-pin).
3.
Use the power cable supplied to connect the supply / evaluation unit to the permissible
supply voltage.
4.
Switch on the measuring system using the master On / Off switch.
Important
Before the device switches to the actual measuring mode, it performs a self-test, during which
all the indicators on the "DISPLAY" light up. After a minute, the device is ready for operation. It
then takes approximately half an hour to reach maximum measuring accuracy.
The measuring rate is set to N = 100. If you require a different measuring rate (option), please
refer to section 5.3.1.
Whenever communication between the flowmeter sensor and supply / evaluation unit is
interrupted, e.g., after the flowmeter sensor has been changed, you must remember to press
the "RESET" button.
If no flowmeter sensor is connected, the "EO2" error message appears to indicate this.
If the "E11" error message appears, it means that you have either pressed the "TEMP" button
when there is no temperature sensor with a temperature option or that an error has occurred
in the temperature evaluation system.