ABB Sensyflow FMT700-P Operating	 Instruction Download Page 24

Communication 

 

24 Sensyflow 

FMT700-P 

42/14-42-EN 

 

Pos: 15.1 /Überschriften/1/J - L/Kommunikation @ 0\mod_1139412848859_3101.doc @ 3153 @ 11 

Kommunikation 

6 Communication 

 

Pos: 15.2 /Überschriften/1.1/1-spaltig/D - F/Digitale Schnittstellen @ 33\mod_1253108530343_3101.doc @ 301241 @ 2 

 

6.1 Digital 

interfaces 

 

Pos: 15.3 /Kommunikation/Durchfluss/FMT700-P/Digitale Schnittstellen @ 33\mod_1253168721390_3101.doc @ 301619 @  

 

The digital interfaces provide the same mass flowrate value as the analog output. They enable 
you to connect the device directly to digital systems without suffering any loss of accuracy as a 
result of an additional conversion process. All the signal lines of the digital interface are 
electrically isolated. 
 

Electrical isolation: 

Electrical isolation is achieved by means of an optocoupler consisting of an LED and a 
phototransistor. 
 

G00941

 

Fig. 19:  Optocoupler 
 

Pos: 15.4 /==== Leeres Modul mit zwei Absatzmarken, DS, 1-spaltig ==== @ 3\mod_1159257408468_0.doc @ 42813 @  

 

 
 

Pos: 15.5 /Kommunikation/Durchfluss/FMT700-P/Serielle Schnittstelle @ 33\mod_1253168930546_3101.doc @ 301667 @ 3 

 

6.1.1 Serial 

interface 

 

The serial interface (V24 / RS 232 C) supports the digital transmission of measured values and 
can also be used to control all the possible operating functions of the supply /°evaluation unit 
(bidirectional communication). 
For the purpose of serial data transmission, only three lines are required for the systems 
involved to exchange all the necessary communication signals. Addresses, data and control 
information are transferred bit by bit. A distinction is made between synchronous and 
asynchronous serial interfaces. 
 

Synchronous serial data transmission 

The transmitter inserts an additional synchronization character into the data stream at defined 
intervals. At the receiver end, this character is detected using special logic. Any shifts are 
corrected to ensure that the transmitter and receiver are synchronized. 
 

Summary of Contents for Sensyflow FMT700-P

Page 1: ...hermal Mass Flowmeter Sensyflow FMT700 P Pos 1 Titelbl tter Copyright BA IA Durchfluss FMT700 P 33 mod_1251794739611_3101 doc 298458 for air test benches Pos 2 Titelbl tter Copyright Copyright Seite 3...

Page 2: ...222 580 Fax 49 621 381 931 29031 automation service de abb com Copyright 2010 by ABB Automation Products GmbH Subject to changes without notice This document is protected by copyright It assists the...

Page 3: ...connection 12 3 2 SLOT 1 D SUB connection Serial Output 12 3 3 SLOT 3 D SUB connection Totalizer 13 3 4 SLOT 4 D SUB connection Analog Outputs 14 4 Commissioning 16 5 Operation 17 5 1 Supply evaluati...

Page 4: ...Contents 4 Sensyflow FMT700 P 42 14 42 EN 10 3 Approvals and certifications 46 10 4 Declaration of conformity 47 11 Index 49...

Page 5: ...uctions and all safety warning symbols in these instructions can optimum protection of both personnel and the environment as well as safe and fault free operation of the device be ensured Information...

Page 6: ...indicates operator tips particularly useful information or important information about the product or its further uses It does not indicate a dangerous or damaging situation Pos 4 11 berschriften 1 1...

Page 7: ...purposes Fill out the return form see the Appendix and include this with the device The EU Directive governing hazardous materials dictates that the owners of any hazardous waste are also responsible...

Page 8: ...2 96 EC or relevant national laws e g ElektroG in Germany The product must be disposed of at a specialist recycling facility Do not use municipal garbage collection points According to the WEEE Direct...

Page 9: ...Fig 2 12 Display 13 Measuring rate 14 Selection buttons TEMP RESET SAMPLE RATE UNIT SELECT SENSOR SELECT For selecting the fluid temperature For performing a master reset Switch for setting the measur...

Page 10: ...note on rear Notice Potential damage to parts Check whether the voltage specified on the name plate matches the line voltage G00932 Fig 4 Pos 6 6 Aufbau und Funktion Durchfluss FMT700 P D SUB Stecker...

Page 11: ...01132 3 2 2 3 BNC outputs and sensor connection ANALOG OUT 0 10 V Analog output voltage via BNC socket ANALOG OUT 0 4 20 mA Analog output current via BNC socket SENSOR 1 Connection for flowmeter senso...

Page 12: ...0 5 Flowmeter sensor coding 21 6 Flowmeter sensor coding 22 7 Flowmeter sensor coding 23 G00924 Fig 7 Terminal connection of BNC socket and flowmeter sensor socket Pos 8 3 Leeres Modul mit einer Absat...

Page 13: ...ssignment 1 20 LSB1 14 213 2 21 15 214 MSB2 3 22 16 4 23 17 5 24 18 6 25 19 7 26 20 8 27 21 9 28 22 10 29 23 REMOTE CTRLSTART STOP 11 210 24 Uext 12 211 25 GNDext 13 212 1 LSB Least Significant Bit 2...

Page 14: ...er coding 19 Nominal diameter coding 7 Free coding 20 Free coding 8 Free coding 21 Free coding 9 22 10 23 11 Temp 1 24 Temp 1 12 Temp 2 25 Temp 2 13 GNDext G00925 25 14 13 1 Fig 11 Terminal connection...

Page 15: ...0 0 0 00 DN 25 1 1 0 0 30 DN 50 1 0 1 0 50 DN 80 1 1 1 0 70 DN 100 1 0 0 1 90 DN 150 1 1 0 1 B0 DN 200 1 0 1 1 D0 Special 1 1 1 1 F0 Pin 4 Pin 17 Flowmeter sensor 1 1 1 Flowmeter sensor 2 1 0 Medium...

Page 16: ...e switches to the actual measuring mode it performs a self test during which all the indicators on the DISPLAY light up After a minute the device is ready for operation It then takes approximately hal...

Page 17: ...an cater for all flowmeter sensor nominal diameters with a standard measuring range It also makes for user friendly operation The device is available as a 19 plug in module or as a portable desktop un...

Page 18: ...figurable number of individual measurements You can set the required number of individual measurements using the selector switch on the front of the supply evaluation unit A measuring cycle period tim...

Page 19: ...ynamic processes If a high measuring rate is set the average is produced from a large number of individual measurements and undesirable fluctuations can be eliminated Pos 13 8 Bedienung Durchfluss FMT...

Page 20: ...sient recorder A process computer A controller This makes it possible to visualize and record rapidly changing flowrates The Sensyflow FMT700 P measuring system has been specially developed for applic...

Page 21: ...ase of the current output Current output 0 20 mA 4 20 mA Max load impedance 550 Max load impedance 550 G00938 FMT700 P Fig 16 Additional units at the current output Voltage output 0 10 V Max current l...

Page 22: ...ndergo downstream electrical processing a jumper must be removed from inside the device by the factory In such cases the electrical circuit outside the device has to be completed using a display unit...

Page 23: ...ly those measured values for the flowmeter sensor selected via the SENSOR SELECT button are shown on the DISPLAY and output as an optional digital and or analog signal The set measuring rate see secti...

Page 24: ...ptocoupler Pos 15 4 Leeres Modul mit zwei Absatzmarken DS 1 spaltig 3 mod_1159257408468_0 doc 42813 Pos 15 5 Kommunikation Durchfluss FMT700 P Serielle Schnittstelle 33 mod_1253168930546_3101 doc 3016...

Page 25: ...V 8 V 3 V 8 V 3 V Fig 20 Voltage level of transmit and receive data Pos 15 7 Kommunikation Durchfluss FMT700 P Datenformat 33 mod_1253169151718_3101 doc 301739 4 6 1 1 2 Data format Data transmission...

Page 26: ...ume in g l once the stop bit has been set externally triggered XZ Measuring rate change for display and analog out Example XZ0100CR XZ0005CR 4 character cycle number must be used XSR Requests the set...

Page 27: ...utput Certain switchover functions are excluded and can only be changed on the the front panel XTT General information about the interface is transmitted see section 6 1 1 5 XTR T 035 4 C XA CLS SU 12...

Page 28: ...te for display 0100 9999 measuring cycles Measuring rate for analog out 0001 9999 measuring cycles Fast mode Display not used each measured value is sent to the analog output Temperature option Measur...

Page 29: ...Notice 4 characters must always be entered when setting the measuring rate see section 5 3 1 e g for a SAMPLE RATE of 50 the command XZ _ _ 50 CR must be entered During constant transmission mode no o...

Page 30: ...perating mode and simultaneously initiates the measuring process The STOP signal marking the end of the measuring process is triggered by a rising edge at the REMOTE CTRL input see section 3 3 Uextern...

Page 31: ...ace Following the STOP signal the mass flowrate value m is made available at the digital output for approx 15 seconds in the form of a digital 15 bit binary signal along with the DATA READY BIT For de...

Page 32: ...COLLECTORS see section 3 3 Permissible measurement period TM 10 s TM 115 s Error due to measurement interval 0 05 with TM 100 s 0 15 with TM 10 s Measurement uncertainty when determining air mass 2 Au...

Page 33: ...measuring fluid must match the arrow on the probe Particles and fibers can impair measuring accuracy We therefore recommend that you use an air filter from our range of accessories for filtering purpo...

Page 34: ...daptation options DN 25 DN 50 DN 80 DN 100 Tapered flanges are centrically compressed using a tension clip vacuum clamping technology DN 150 DN 200 Tapered flanges are centrically compressed using a t...

Page 35: ...lue Repeatability 0 25 of measured value Influences Temperature effect 0 03 K of measured value Pressure effect 0 2 100 kPa bar of measured value Response time T63 12 ms Pressure drop flowmeter sensor...

Page 36: ...Pos 24 Leeres Modul mit einer Absatzmarke DS 1 spaltig 2 mod_1153381574375_0 doc 35553 Pos 25 Leeres Modul mit einer Absatzmarke DS 1 spaltig 2 mod_1153381574375_0 doc 35553 Pos 26 Technische Daten D...

Page 37: ...3101 doc 294111 2 9 1 Flowmeter sensor Sensyflow FMT700 P DN 25 Pos 28 3 Technische Daten Datenblatt Durchfluss FMT700 P Ma zeichnungen Abmessungen FMT700 P DN 25 32 mod_1249559479784_3101 doc 291961...

Page 38: ...200 33 mod_1250757701399_3101 doc 294183 L L2 L1 L2 di D D1 B H G00904 Fig 27 Dimensions in mm inch DN D D1 di L L1 L2 B H 50 64 2 52 80 0 3 15 58 2 28 184 7 24 29 5 1 16 88 0 3 46 80 89 3 50 108 5 4...

Page 39: ...Technische Daten Datenblatt Durchfluss FMT700 P Ma zeichnungen Zubeh r 33 mod_1251963704087_3101 doc 299181 In order to simplify the installation of our measuring system for the user we recommend to...

Page 40: ...lange on one side KF KF KF KF KF KF appr 244 9 61 269 10 59 144 5 67 34 1 34 2 5 0 10 2 5 0 10 2 5 0 10 2 5 0 10 2 5 0 10 2 5 0 10 appr 110 D2 D2 4 33 D2 D Di D2 D2 D Di D2 D4 D3 10 x D inlet section...

Page 41: ...Inner center ring Clamping ring FL Optimal AS 15 0 59 57 2 24 46 1 81 16 0 63 55 2 17 73 2 87 4 0 16 Sectional detail pipe connection without clamping ring KF KF 24 0 94 27 1 06 4 0 16 10 0 39 10 0 3...

Page 42: ...nge on one side 5 x D outlet section with flange on one side KOF KOF KOF L6 L7 25 0 98 2 0 08 2 0 08 25 0 98 Di D2 D D2 D2 D Di D2 2 0 08 D6 D2 L5 D7 Standard measuring section Measuring section 3 Flo...

Page 43: ...2 ZWF D2 2 0 08 2 0 08 D2 2 0 08 KOF 4 0 16 Sectional detail pipe connection without clamping ring 2 0 08 Di D2 D2 2 0 08 2 0 08 Individual planning G00920 ZWF L9 ZWF Fig 33 Dimensions in mm inch DN L...

Page 44: ...D2 D2 D Di D2 D2 D Di D2 Di D Flow conditioner with flange 10 x D inlet section with flange on one side 5 x D outlet section with flange on one side KOF KOF KOF S S S L6 L7 4 0 16 4 0 16 Di D2 D2 Di...

Page 45: ...O ring Tapered flange ZWF KOF D2 4 0 16 D2 4 0 16 37 1 46 L1 25 0 98 25 0 98 Sectional detail pipe connection without clamping ring Di 1 0 04 4 0 16 D Individual planning G00921 ZWF L9 ZWF Fig 35 Dim...

Page 46: ...layer of polythene that is 0 2 mm thick Use an amount of desiccant that is appropriate for the packing volume and the expected transport time at least sufficient for 3 months You should also line the...

Page 47: ...L Konformit tserkl rung 29 mod_1242206940830_3101 doc 270646 2 10 4 Declaration of conformity Pos 30 10 Konformit tserkl rungen Zertifikate Durchfluss FMT700 P EG Konformit tserkl rung 39 mod_1273484...

Page 48: ...tact person Telephone Fax E mail Device details Type Serial no Reason for the return description of the defect Was this device used in conjunction with substances which pose a threat or risk to health...

Page 49: ...FMT700 P DN 50 DN 200 38 Front view 9 G General information 28 H Hazardous materials 7 How the totalized value is represented 31 I Information on WEEE Directive 2002 96 EC 8 Integrated management sys...

Page 50: ...ined herein without notice Printed in the Fed Rep of Germany 05 2010 ABB 2010 3KXF421006R4201 42 14 42 EN Rev C ABB Ltd Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel 44 0 1946 830 611 Fax 44...

Reviews: