ABB PGC5000 Series Service Instruction Download Page 11

 

 

892J009MNAE |  DBDID  | 

11

 

 

3.4

 

Symptom – Excessive Noise 

Possible Cause 

Action(s) 

Bad electrical connections 

Check the signal cable connection and the bias voltage cable 

connection.   

Check the signal cable shield for fit. The fitting should be snug. 

Verify connection between ground rod and ground. 

Air Leak 

Leak-check all of the fittings within the oven system to verify a leak-

free system. 

Insufficient clearance for 

the plasma electrode  

The plasma O-ring on the plasma tube is designed to prevent this 

issue. If the plasma electrode appears to be too close to the 

Swagelok

®

 fitting, remove power from the analyzer and remove the 

plasma tube from the top nut. 

Install or replace the plasma O-ring on the plasma tube and re-install 

the plasma tube. 

 

3.5

 

Symptom – Elevated baseline 

Possible Cause 

Action(s) 

Poor carrier gas 

Background signal is directly linked to the quality of gases being used. 

UHP (ultra-high purity) gases should always be used. If gas quality is 

suspected, replace gas with new UHP quality gas. 

When UHP gases cannot be used, a scrubber will need to be installed 

to remove excess levels of oxygen, water, and trace hydrocarbons. 

Air Leak 

An air leak in the chromatographic system can also cause an elevated 

baseline. Verify that there are no air leaks in the chromatographic 

system. 

 

3.6

 

Symptom – Baseline drift 

Possible Cause 

Action(s) 

Poor pressure control 

Pressure controls not providing a consistent, pulse free, flow rate to 

the detector can cause detector drift. 

Monitor the pressure gauges on the supply cylinder and measure the 

flow out of the exhaust port over a one-hour time period to insure 

correct flow conditions. 

Column bleed 

Re-condition or replace the column. 

Plasma cell leak 

Initially, this may appear as column bleed, but the signal will 

equilibrate as the oven reaches an isothermal temperature. 

Reset the plasma cell in the Swagelok

®

 fitting and tighten to leak-free. 

Contamination or late-

eluting peaks 

It is possible that there is a late-eluting peak or contamination that is 

coming off the column several runs after it was introduced. 

Alter the cycle time of the injection to see is the baseline drift changes 

location. 

If this is confirmed, install filters to eliminate the contamination. 

Extend the instrument cycle time to elute the material prior to the 

next analysis. 

 

3.7

 

Symptom – Spikes 

Possible Cause 

Action(s) 

Loose connections 

Verify that the signal cable is securely attached to the electrometer 

and the BNC fitting on the collector shield. 

Missing ground leads 

Verify that the ground lead from the plasma center electrode is 

connected. If it is not, power the analyzer down and re-connect the 

center electrode ground lead. 

Verify that the main power supply ground lead is connected to the 

appropriate oven chassis ground point. 

Summary of Contents for PGC5000 Series

Page 1: ...ABB MEASUREMENT ANALYTICS 892J009MNAE PGC5000 Gas Chromatograph Dielectric Barrier Discharge Ionization Detector DBDID SERVICE INSTRUCTION Measurement made easy...

Page 2: ...Symptom Spikes 11 4 Specifications 12 4 1 Physical 12 4 2 Power 12 4 3 Modes of operation 12 4 4 Detector sensitivity 12 4 5 Gas consumption 12 5 Drawings 12 List of Figures Figure 1 1 Dielectric Bar...

Page 3: ...d safety regulations concerning the construction and operation of electrical installations The regulation on technical working materials safety guidelines for tools The regulations and recommendations...

Page 4: ...DBDID uses a high voltage alternating current discharge to create a plasma The system operates at a high voltage and a low current which is applied to the glass reactor tube at the round metal electr...

Page 5: ...high voltage alternating current applied to a gas as it flows through a dielectric material The gas is typically helium or argon and the dielectric material is a quartz or Pyrex tube Two electrodes a...

Page 6: ...injury High voltage present Make sure that power to the PGC5000 has been disconnected before removing the detector To remove the detector from the PGC5000 perform the following steps Disable the isoth...

Page 7: ...t the coax cable item 1 in Figure 2 3 from the DBDID amplifier assembly Figure 2 3 DBDID amplifier assembly 1 On the DBDID power supply board unplug the high voltage power connector J2 item 3 the bias...

Page 8: ...g the correct flow 1 Set flows by measuring the flow at the cell vent tube Refer to the data pack for correct flows The detector should automatically light when power is applied To verify remove the f...

Page 9: ...ee the last page of this instruction for contact information Always include the information listed in Equipment Identifications and Configuration Identification in your request When ordering parts for...

Page 10: ...a tube in a dark environment There should be a pink or almost white glow in the plasma tube If the tube is blue and the signal is low there is nitrogen in the plasma Verify that the correct reaction g...

Page 11: ...system 3 6 Symptom Baseline drift Possible Cause Action s Poor pressure control Pressure controls not providing a consistent pulse free flow rate to the detector can cause detector drift Monitor the p...

Page 12: ...ed Argon ionization mode AID In this mode the detector is not sensitive to fixed gases and methane but is sensitive to a wide range of organic and inorganic volatile components It is capable of ionizi...

Page 13: ...to optional purge system 2 Reaction gas inlet plasma gas 3 Purge gas vent Detector assembly purge not used in most applications Flameproof breather must remain open or connected to optional purge syst...

Page 14: ...14 DBDID 892J009MNAE Figure 5 3 Diagram flow PGC5000 air and carrier...

Page 15: ...892J009MNAE DBDID 15 Figure 5 4 PGC5000 oven electronics internal connections diagram sheet 1...

Page 16: ...16 DBDID 892J009MNAE Figure 5 5 PGC5000 oven electronics internal connections diagram sheet 2...

Page 17: ...echnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for...

Reviews: