ABB PGC5000 Series Service Instruction Download Page 10

10

 | DBDID |  892J009MNAE

 

To determine the source of the problem:  

1.

 

Identify whether the problem is with the chromatographic oven or the detector. In general, if 

the problem is with component retention times and not component sensitivity, the problem is in 

the chromatographic oven and not the detector. 

 

Ask: “What was the last thing done to the system?” Most problems are associated with the 

latest change made to the system, so this is the best place to begin look for the problem. 

If the considerations above identify the detector as the primary cause of the malfunction, use the 

following sections to help identify and correct the problem. 

3.1

 

Symptom – Plasma won’t light 

Possible Cause 

Action(s) 

No power 

If the fuse is intact, contact customer support for further assistance. 

Air in plasma 

The presence of too much air in the plasma will often cause the 

plasma not to light. 

Purge the system with reaction gas for at least 15 minutes. 

If the plasma still will not light, check for air leaks in all of the fittings. 

Back pressure too high  

If there is too high of a restriction in the exhaust line, the detector 

plasma will not light. 

Replace the exhaust/signal tube with a new one and see if the plasma 

will light. 

Wrong plasma gas 

Verify that the reaction gas is either helium or argon. Replace if 

necessary. 

Defective electronics 

See sections 

3.2

 and 

3.3

. 

 

3.2

 

Symptom – No signal 

Possible Cause 

Action(s) 

No Power 

If the fuse is intact, contact customer support for further assistance. 

Air in plasma 

Observe the plasma tube in a dark environment. There should be a 

pink or almost white glow in the plasma tube. If the tube is blue and 

the signal is low, there is nitrogen in the plasma. 

Verify that the correct reaction gas (helium or argon) is being used. 

Turn off the power to the plasma and check the gas supply, check for 

leaks, and verify the proper flow rate. 

Signal cable problem  

Verify that the collector shield BNC fitting contacts the exhaust tube. 

To do this, remove the BNC cable from the collector shield. Verify 

contact between the center pin of the BNC and the exhaust tube using 

a VOM or continuity tester. Re-connect the BNC cable to the collector 

shield. Verify the cable is connected to the collector shield and the 

electrometer board. 

Wrong bias voltage 

Using a VOM meter, verify the voltage between the Bias test point and 

chassis ground on the Power Supply Board. The voltage should be +10 

volts DC. 

 

3.3

 

Symptom – Low signal 

Possible Cause 

Action(s) 

Leakage in the 

chromatographic system 

Verify retention times for the analytes of interest.   

If OK, check for leakage on all the fittings on the detector end of the 

system. 

If retention times have shifted, check for leaks on the sample 

introduction side of the system. 

Wrong reaction gas 

Using nitrogen as a reaction gas will result in a low signal. Verify that 

the proper reaction gas is being used. Replace if necessary. 

Plugged exhaust tube  

Unique problem that will manifest itself with a lower signal and slower 

retention times. 

Visually inspect the exhaust tube to identify if any materials have 

plugged the tube. If plugged, insert a cleaning wire into the exhaust 

tube to remove the blockage.   

If these actions do not correct the problem, replace the exhaust tube. 

Summary of Contents for PGC5000 Series

Page 1: ...ABB MEASUREMENT ANALYTICS 892J009MNAE PGC5000 Gas Chromatograph Dielectric Barrier Discharge Ionization Detector DBDID SERVICE INSTRUCTION Measurement made easy...

Page 2: ...Symptom Spikes 11 4 Specifications 12 4 1 Physical 12 4 2 Power 12 4 3 Modes of operation 12 4 4 Detector sensitivity 12 4 5 Gas consumption 12 5 Drawings 12 List of Figures Figure 1 1 Dielectric Bar...

Page 3: ...d safety regulations concerning the construction and operation of electrical installations The regulation on technical working materials safety guidelines for tools The regulations and recommendations...

Page 4: ...DBDID uses a high voltage alternating current discharge to create a plasma The system operates at a high voltage and a low current which is applied to the glass reactor tube at the round metal electr...

Page 5: ...high voltage alternating current applied to a gas as it flows through a dielectric material The gas is typically helium or argon and the dielectric material is a quartz or Pyrex tube Two electrodes a...

Page 6: ...injury High voltage present Make sure that power to the PGC5000 has been disconnected before removing the detector To remove the detector from the PGC5000 perform the following steps Disable the isoth...

Page 7: ...t the coax cable item 1 in Figure 2 3 from the DBDID amplifier assembly Figure 2 3 DBDID amplifier assembly 1 On the DBDID power supply board unplug the high voltage power connector J2 item 3 the bias...

Page 8: ...g the correct flow 1 Set flows by measuring the flow at the cell vent tube Refer to the data pack for correct flows The detector should automatically light when power is applied To verify remove the f...

Page 9: ...ee the last page of this instruction for contact information Always include the information listed in Equipment Identifications and Configuration Identification in your request When ordering parts for...

Page 10: ...a tube in a dark environment There should be a pink or almost white glow in the plasma tube If the tube is blue and the signal is low there is nitrogen in the plasma Verify that the correct reaction g...

Page 11: ...system 3 6 Symptom Baseline drift Possible Cause Action s Poor pressure control Pressure controls not providing a consistent pulse free flow rate to the detector can cause detector drift Monitor the p...

Page 12: ...ed Argon ionization mode AID In this mode the detector is not sensitive to fixed gases and methane but is sensitive to a wide range of organic and inorganic volatile components It is capable of ionizi...

Page 13: ...to optional purge system 2 Reaction gas inlet plasma gas 3 Purge gas vent Detector assembly purge not used in most applications Flameproof breather must remain open or connected to optional purge syst...

Page 14: ...14 DBDID 892J009MNAE Figure 5 3 Diagram flow PGC5000 air and carrier...

Page 15: ...892J009MNAE DBDID 15 Figure 5 4 PGC5000 oven electronics internal connections diagram sheet 1...

Page 16: ...16 DBDID 892J009MNAE Figure 5 5 PGC5000 oven electronics internal connections diagram sheet 2...

Page 17: ...echnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for...

Reviews: