![ABB Navigator 600 Series User Manual Download Page 8](http://html.mh-extra.com/html/abb/navigator-600-series/navigator-600-series_user-manual_4467008.webp)
Navigator 600 Series Analyzers
Silica and Phosphate
Appendix A – GSD File Module
6
IM/NAV6PBS
Issue 1
Appendix A – GSD File Module
A.1 Module 01
Table A.1 defines the Module 01 data available from a Navigator
600 Series analyzer via Profibus cyclic transfer:
A.2 Status Byte Definition
Table A.2 defines the meaning of the contents of a Navigator
600 Series analyzer's status byte.
Note.
Float = Floating-point number – requires 4 bytes
Char = Character – requires 1 byte
Int
= Integer – requires 2 bytes
Byte
Type
Description
1 to 4
Float
Stream 1 value
5
Char
Stream 1 status
6 to 9
Float
Stream 2 value
10
Char
Stream 2 status
11 to 14
Float
Stream 3 value
15
Char
Stream 3 status
16 to 19
Float
Stream 4 value
20
Char
Stream 4 status
21 to 24
Float
Stream 5 value
25
Char
Stream 5 status
26 to 29
Float
Stream 6 value
30
Char
Stream 6 status
Table A.1 Module 01 Data
Byte (Hex)
Definition
1C
OOS (Out of Service)
80
Good
89
Low alarm limit advisory alarm
8A
High alarm limit advisory alarm
8D
Low-low alarm limit critical alarm
8E
High-high alarm limit critical alarm
Table A.2 Status Byte Definition