Navigator 600 Phosphate
Multi-stream
5 Configuration
IM/NAV6P/MS–EN Rev. F
31
5.4 Measurement
There are three tabs in the 'Measurement' screen:
5.4.1 Setup
Fields
Description
Display Units
Two sub-parameters are available at the
Display Units
parameter:
Display Units
The display units for phosphate concentration – 3 options are available:
ppm
mg/l
mg/kg
Chemical Units
The chemical units displayed – 2 options are available:
Phosphate PO4
Phosphorus P
Control Temp
The reaction temperature range is 37 to 50 °C (98 to 122 °F) in 0.1 °C (0.18 °F) increments. The
analyzer is optimized to run at a reaction temperature of 45 °C (113 °F).
Any changes to the reaction temperature may affect analyzer performance.
Any changes to the pre-heater temperature may affect analyzer performance.
Enable Streams
The stream(s) to be enabled for sampling.
Stream Sequence
The order in which the streams are sampled.
Up Scale Time
The time taken for the analyzer to measure a full step change from a low concentration sample to a
high concentration sample (14 to 60 minutes).
Down Scale Time
The time taken for the analyzer to measure a full step change from a high concentration sample to a
low concentration sample (14 to 60 minutes).