Navigator 600 Phosphate
Multi-stream
4 Installation
16
IM/NAV6P/MS–EN Rev. F
4.5.2 Mounting the Analyzer
Referring to Fig. 4.6:
1. Mark the wall using the dimensions shown or, with the
analyzer supported carefully against the wall, mark
through the mounting holes.
2. Drill and plug 3 holes
A
and
B
, suitable for M6 or
1
/
4
in.
fixings.
3. Screw in top fixing
A
, leaving a gap of 20 mm
(0.78 in.) between the fixing head and the wall.
4. Hang the analyzer onto fixing
A
, ensuring the analyzer is
retained firmly against the wall.
5. Secure the analyzer to the wall using 2 fixings
B
.
4.5.3 Mounting the Reagent Shelf (Optional)
If used, mount the reagent shelf no more than 1 m (3.3 ft.) from
the analyzer’s bottom plate – see Fig. 4.5, page 15. The shelf
can be attached through the back plate or either of the side
plates.
Referring to Fig. 4.7:
1. Mark the wall using the dimensions shown or, with the
shelf supported carefully against the wall, mark through
the mounting holes.
2. Drill and plug holes suitable for M8 or
5
/
16
in. fixings.
3. Hang the shelf onto the screws and tighten the fixings to
secure the shelf firmly against the wall.
Dimensions in mm (in.)
Fig. 4.6 Mounting the Analyzer
Note.
Clearance – the enclosure doors can open 180°. If
mounting in a confined area, allow sufficient clearance for
cables on the door hinge side (min. 270 mm, 10.6 in.) and
100 mm (3.93 in.) on door opening side.
Note.
It is not possible to adjust fixing
A
once the
analyzer is placed over it. If necessary, remove the
analyzer and adjust the fixing.
525
(20.7)
200
(7.8)
All Dimensions in mm (in.)
Fig. 4.7 Reagent Mounting Shelf (option)
81
(3.18)
165
(6.59)
82
(3.23)
75