CO M PAC T U LT R A S O N I C L E V EL T R A N SM I T T ER | L S T 3 0 0 – EN R E V. E
43
Example tank
In this example, a cylindrical tank with a cone base is
configured . This is a simple shape to configure, containing only
3 points and linear sections in between .
Figure 26 Cylindrical tank with cone base
Figure 27 Linearization curve of cylindrical tanks with cone base
Configure the level settings
1
Log on to the LST300 at the
Standard
or
Advanced
level .
2
Select
Easy Setup
from the main menu .
3
Select
Volume
as the
Operate Mode
and then press
Next
.
4
Select Empty Distance to configure the distance from the
face of the sensor to the bottom of the tank .
In Level mode, Empty Distance indicates the tank is empty . In
this example the sensor is assumed as 0 .25 m above the top of
the tank .
5
Press
Next
to confirm the Empty Distance configuration .
6
Configure
Span
to configure the distance from the bottom
of the tank to the top of the tank .
In Level mode, Span indicates the tank full position . This is the
3 .00 m
as shown in figure 22 “Cylindrical tank with cone base” .
7
Press
Next
. to confirm the Span configuration .