USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
29
Connect the GLA351-CCIA3 analyzer to the pretreatment box
1.
Connect the provided 6’ x 3/8” tubing from the external pump to the pretreatment
box TO EXT. PUMP port.
2.
Connect the pretreatment box from the ANALYZER EXIT port to the analyzer’s TO
EXT. PUMP port, using the provided 6’ x 3/8” Swagelok connector and tubing.
3. Connect the pretreatment box ZERO AIR OUT port to the ZERO AIR INLET port on
the analyzer, using the provided 6’ x 1/4” Swagelok connector and tubing.
4. Connect the pretreatment box TO ANALYZER port to the INLET port on the
analyzer, using the provided 6’ x 1/4” Swagelok connector and tubing.
5. Attach the pretreatment box OPEN port to the external Soda Lime / Litholyme
canister, using the 3’x1/4” Swagelok connector and tubing.
a. Leave the opposite end of the canister open to the atmosphere.
6. Connect the provided exhaust muffler to the exhaust port of the external pump to
expel into the room air, or route to your facility ventilation system. (Figure 15)
Figure 16: Exhaust Muffler
7.
Connect a ¼” sample tube (not provided) from the SAMPLE IN port on the
pretreatment box to your sample source.
Figure 17 shows the pretreatment box plumbing diagram.
Figure 17: Pretreatment Box Plumbing Diagram