Troubleshooting table
In the following table, a series of typical situation are indicated, useful to understand and solve faults or improper working
Note 7: In case none of the following cases will help you to find a solution, please contact the nearest ABB service
organization
Nr. Situation
Description
Suggestions
1
TS3 does not power on
The display does not switch on
1.
Check the power cable is properly
connected.
2.
Check that the power switch is in on
position
3.
Check the presence of the fuse on fuse-
holder, and that the fuse is not interrupted.
“Output on” led does not switch on
1.
Press Start/ Stop button
Message “NO LOAD” on the
display
Check that the test cable and extension are
correctly connected to trip unit
2
TS3 does not generate
current
1.
Message “OVERVOLTAGE”
on the display
2.
Message “UNDERVOLTAGE”
on the display
3.
Message “OVERHEAD” on the
display
1.
Network voltage is over functional limits
2.
Network voltage is below functional limits
3.
Internal temperature is over operational
limits
3
TS3 does not stop the
current generation
During the test of a Trip unit the
TS3 does interrupt the current
generation
1.
Check if DUT and TS3 settings are correct.
2.
Check if the Itest current is higher that the
current threshold adjusted on the DUT,
considering the tolerances
3.
Check the connections of testing cable
4
Unexpected current
interruption
During the test of a Trip unit the
TS3 unexpectedly interrupt the
current generation
1.
Check the connections of testing cable
(verify that the message “NO LOAD” is not
present on TS3 display)
5
Tripping times is longer
or shorter that expected
TS3 stop the current generation
earlier or later that expected
1.
Check that the Itest current and adjusted
paramenters on DUT are correct
2.
Check that the output signal frequency is
the same of what is adjusted on the DUT
6
“Wrong Password”
Message “Wrong password” on the
display
Check the exactitude of the inserted password
7
Forgotten the Password
In case of lost personalized
password
Please contact the nearest ABB service
organization
8
The FW version of TS3
is 1.00
In case of the FW version is 1.00
and missing the communication
between the TS3 and PC
Install the TS3 Manager and perform the
Firmware upgrade.
Only from FW version 2.00 on it is possible to
establish the communication.
9
Communication problem In case of missing communication
between PC and TS3
1.
Check if the USB cable is connected
between TS3 and PC.
2.
Install the driver USB-RS485 converter.
3.
Check which port COM is used.
4.
Check which version of FW is installed on
TS3; only from version 2.00 on is possible
to use the TS3Manager.
5.
Check the ID number selected from TS3
Manager ( the default number is 1).