USER MANUAL
| ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
53
8.
Each step is displayed in the lower-right panel of the calibration screen as the
analyzer performs the calibration. Figure 38 shows the calibration process as a flow
chart.
Figure 38: Calibration Flow
9.
When the
Calibration Complete
message is displayed, click the
CLOSE
button.
10.
Enter
TimeChart
by selecting the
Display
button on the
User Interface Control Bar
,
and verify that the displayed concentration correctly corresponds to your local gas
standard.
11.
Repeat steps 1-10 for each gas you wish to calibrate.
The time of latest calibration is stored in
Reference Gas Settings
within
the
Calibration
menu for future reference.