… 13 Maintenance
Instrument information
Menu Path
‘
Maintenance /
Diagnosis /
Device Info’
Figure 115: ‘Device information’ Menu
Display
Both the serial number and the version number, the date of the
software and the hardware index are shown for the device and
the integrated assemblies.
Operate pump
Menu Path
‘
Operation /
Pump’
Figure 116: ‘Pump settings’ menu
It is also possible in the configurator to power-up and power-
down the pump and to change the pump power.
Pump as a component of the ‘Integral gas feed’ option
A pump can be installed in the model EL3020 gas analyzer as a
component of the ‘Integral gas feed’ option (not available for
Limas23).
Note
In this case, a flow sensor is also installed.
Power-up and power-down of the pump
The pump can be powered-up and powered-down manually,
for example for maintenance reasons or in an emergency. The
emergency shutdown cannot be overridden by the automatic
calibration.
Setting the pump power
The pump power can be changed manually when the pump is
switched on. The measured value of the integrated flow sensor
is displayed for setting the pump power.
Note
The pump cannot be operated while an automatic calibration is
in progress.