Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
6 Configuration
32
IM/AZT6CR–EN Rev. J
6.2 Measurement
There are three tabs in the Measurement screen:
6.2.1 Setup
6.2.2 Streams
6.2.3 Clean
Fields
Description
Measurement Rate
Sets the number of samples analyzed per hour.
Chemical Units
For certain parameters there is a choice of units to display the results.
Measuring Units
The results can be expressed in a variety of units, for example by weight (mg or µg) or by volume
(ppm or ppb).
Temperature Units
The results can be expressed in either Celsius (°C) or Fahrenheit (°F).
Cell Temperature
The cell is temperature controlled and can be set at temperatures between 25 °C (77 °F) and 50 °C
(122 °F).
Stream Sequence
In multi-stream analyzers the streams are measured one after the other. This facility allows the
operator to set the stream sequence, for example stream 1,2,1,2,3 or 1,2,3,3,1,1 etc.
Cell Rinse Sequences
Number of times the cell is rinsed with sample prior to measurement. Configurable between 1 and 4
rinses.
Fields
Description
Stream 1 (2 & 3)
The Stream tag text is displayed in the operator views. Up to 20 characters can be used.
The Stream tag is also displayed in the configuration files and audit log files.
Stream 1 (2 & 3)
Dilution Ratio
Sets the 'Max. Dilution Factor' values for each stream.
Notes
.
Changing the Dilution Ratio affects the analyzer range. For example:if the analyzer range is 0 to
5 ppm, changing the dilution ratio from the default of 1:5 to 1:1 reduces the analyzer range to 0
to 2 ppm.
The Aztec 600 Manganese Low Range does not operate in dilution mode.
Fields
Description
Cleaning Mode
Enables and sets the automatic cleaning mode. This can be configured to occur at calibration or
during measurement.
The following fields are displayed only if 'Cleaning Mode' is set to 'Measure' or 'Calibrate'.
Port
Sets the port the cleaning solution is attached to
Cell/Sample Lines
User can select whether to clean the measuring cell and sample lines or the measuring cell only.
The following field is displayed only if 'Cleaning Mode' is set to 'Measure'.
Cleaning Frequency
Sets the hourly frequency of the cleaning routine (1 to 24).