![ABB AX411 User Manual Download Page 40](http://html.mh-extra.com/html/abb/ax411/ax411_user-manual_4343040.webp)
38
…6
INSTALLATION
…6.5
Panel-mount Analyzer Connections
6.5.2
Connections – Fig. 6.11
Fig. 6.11 Connections, Panel-mount Analyzers
Terminal block A
L
Line
N
Neutral
E
Earth (Ground)
A4
C
A5
NC
Relay 1
A6
NO
A7
C
A8
NC
Relay 2
A9
NO
A10 C
A11 NC
Relay 3
A12 NO
A13 +
Analog Output 1
A14 —
A15 +
Analog Output 2
A16 —
Terminal Block C
(Analog Option Board)
C1
C2
C3
C4
C5
C6
Earth (Ground)
C7
C
C8
NC
Relay 4
C9
NO
C10 C
C11 NC
Relay 5
C12 NO
C13 +
Analog Output 3
C14 —
C15 +
Analog Output 4
C16 —
Temperature Compensator
Connections
Terminal Block B
Common
Third Lead
TC
B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
Common
Third Lead
TC
Temperature Compensator
Connections
* Dual Conductivity Only
** When a 2-wire Pt100, Pt1000 or 'balco 3k' temperature compensator is fitted.
TC = Temperature Compensator.
Terminal Block B
Conductivity Cell Connections
Sensor B*
Sensor A
1
9
TC Common, Link B1 to B2 (Dual Conductivity Only) / B9 to B10**
2
10
TC Third Lead
3
11
TC
4
12
Screen
5
13
Cell (Cell Electrode)
6
14
Cell (Earth Electrode)
7
15
Not Used
8
16
Not Used
Summary of Contents for AX411
Page 1: ...AX400 Series User Guide Model AX411 Dual Input Conductivity Analyzer ...
Page 53: ...51 NOTES ...
Page 54: ...52 NOTES ...