36
Fig. 6.9 Connections, Wall-/Pipe-mount Analyzer
Terminal block A
Terminal Block B
Terminal Block C
(Analog Option Board)
C1
C2
C3
C4
C5
C6
Earth (Gr
ound)
C7
C
C8
NC
Relay 4
C9
NO
C10
C
C11
NC
Relay 5
C12
NO
C13
+
Analog Output 3
C14
—
C15
+
Analog Output 4
C16
—
Temperature Compensator
Connections
Line
L
Neutral
N
Earth (Gr
ound)
E
CA
4
Relay 1
N
C
A
5
NO
A6
CA
7
Relay 2
N
C
A
8
NO
A9
C
A10
Relay 3
N
C
A11
NO
A12
Analog Output 1
+
A13
—
A14
Analog Output 2
+
A15
—
A16
B16
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
TC
Thir
d Lead
Common
TC
Thir
d Lead
Common
Temperature Compensator
Connections
* Dual Conductivity Only
** When a 2-wire Pt100, Pt1000 or 'balco 3k' temperature compensator is fitted.
TC = Temperature Compensator.
Terminal Block B
Conductivity Cell Connections
Sensor B*
Sensor A
1
9
TC Common, Link B1 to B2 (Dual Conductivity Only) / B9 to B10**
2
10
TC Third Lead
3
11
TC
4
12
Screen
5
13
Cell (Cell Electrode)
6
14
Cell (Earth Electrode)
7
15
Not Used
8
16
Not Used
…6
INSTALLATION
…6.4
Wall-/Pipe-mount Analyzer Connections
6.4.2
Connections – Fig. 6.9