AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | O I/AW T4 2 0 - EN R E V. B
61
6
Use the key to edit the buffer value.
Use the
/
keys to set the value and press
the key to confirm changes.
Press the key to proceed to the next step.
7
Place the sensor into buffer 1 and press the
key to perform the calibration. The calibration
process screen is displayed.
On completion the result screen is displayed.
• If the calibration passes, the slope and offset values
are displayed.
• If the calibration fails, the failure reason is displayed.
Note
. The calibration can be canceled at any time during
the process by pressing the
Abort
key.
Calibrate
Abort
Continue
Immerse in Buffer 1
4.0 pH
Calibrate
Abort
pH 4.00 pH
Tmp 25.0 °C
Settling-Please Wait